This paper describes a method for the determination of gross -activity (actinides) in urine. The method consists of a coprecipitation of actinides as Fe/Ca phosphates and hydroxides, a purification as iron phosphate and a LaF3 source preparation; -spectrometry is then carried out without any electroplating procedure. The average recoveries of the method for actinides (except uranium) were about 83.8±8.1%. With a counting time of 1440 min and a counting efficiency of 38% the detection limit was 0.25 mBq/l (6.8 fCi/l). The gross -activity in urine blanks ranged from 0.35 to 0.80 mBq/l (9.4 to 22 fCi/l).