Abstract  

A large area beta scintillation detector has been developed which is currently capable of determining Sr-90/Y-90 contamination in surficial soils. The detector system employs scintillating fiber optic arrays, with active dimensions approximately 15 cm wide by 100 cm long, both ends of which are coupled to multiple photomultiplier tubes (PMTs). Electronic processing includes coincidence requirements to optimize sensitivity and selectivity for the 2.28 MeV (maximum) beta particle from Y-90. Low energy beta particles and gamma rays are discriminated against using double ended and multi-layer coincidence requirements. The detector system is personal-computer-software controlled and data restored in a format compatible with standard database software for ease of final data reduction. Experimental calibration studies have shown a linear response for Sr-90/Y-90 soil concentrations from 12 to over 500 pCi/g and a discrimination factor of 50 to 1 versus Cs-137.