Celeste-1 is a lab-scale hot cell intended for R&D work in reprocessing of low burn-up spent fuel elements. The studies are concerned with head-end, first separation cycle by Purex Process using mini mixer-settlers and development of analytical techniques. The analytical monitoring for process control purposes is based on several off-line techniques, such as X-ray fluorescence spectrometry, potentiometric titration, -and -spectroscopy, spectrophotometry, fluorimetry, density measurement and gas chromatography. The analytical treatment takes place in a shielded working place analytical hot cell, glove boxes and hoods and some final measurements are made in the associated analytical laboratory. A pneumatic system is used for transporting analytical samples. All analytical procedures are ready and in operation.