Abstract  

The alanine dosimeter made for evaluation by diffuse light reflection spectrophotometry (ALA/DRS) does not show the effect of orientation of crystals. Supposed deviation from random orientation has been investigated by EPR spectroscopy. EPR investigation shows that in spite of the very fine size of L-alanine crystals, they are oriented in thin layers of the polyethylene matrix. Specially prepared films with deliberately well oriented crystals have confirmed this observation. Our ALA/DRS dosimeter can be evaluated by the EPR method for the concentration of free radicals, providing that the dominating crystal orientation in the dosimetric film is indicated on it as an arrow, and the sample is inserted into the magnetic cavity always in the same orientation as has been done during the calibration operation.