Author: P. Bode 1
View More View Less
  • 1 Delft University of Technology Interfaculty Reactor Institute Mekelweg 15 2629JB Delft The Netherlands
Restricted access

Abstract  

Detection limit improvement factors have been calculated to predict the improvement in detection limits when replacing a traditional 20% coaxial detector by either a 100% coaxial detector, a well-type detector or by equipping the detector with a Compton suppression shield. It is show that, when measuring sources on the end-cap of the detector, a system with a Compton suppression shield and a 100% detector have almost equal potentials. But when e.g. geometrical consideration larger source-to-end-cap distances are required with coaxial detectors, the well-type detector offers the best outlook for attaining better detection limits. The calculated predictions have been compared with experimental observations and the results agree well.