This work presents a low-level background gamma spectrometry system capable of achieving rapid determination of90Sr without chemical separation, in a soil sample when the90Sr to -ray emitting nuclide concentration ratio exceeds 10. The system consists of a highly efficient HPGe central detector operated in coincidence or anticoincidence with a NaI(Tl) guard detector and a NaI(Tl) backscatter detector. Experimental results indicate that the bremstrahlung radiation from a pure -emitter can be distinguished with respect to the differences in normal and coincidence spectra. The90Sr activity can be directly detemined in the sample in less than 1 hour with a minimum detectable concentration (MDC) estimated at 0.002 Bq·g–1 for a sample containing no appreciable -ray nuclides or other high energy, -emitting nuclides. In actual measurement, a 0.06 Bq·g–1 MDC for contaminated soils is achieved.