Author:
L. Zikovsky Ecole Polytechnique succ. centre-ville C.P. 6079 H3C 3A7 Montréal Québec Canada

Search for other papers by L. Zikovsky in
Current site
Google Scholar
PubMed
Close
Restricted access

Abstract  

Several two and three parameter analytical functions were fitted to the measured detection efficiencies () of a Ge detector for a Marinelli beaker filled with samples up to different heights (H). From these, the expression =308E–0.82/((H–6.1)2+107), gave the best fit. The measured and calculated efficiencies were compared and for 70 determinations, a relative standard deviation of only 7.6% was obtained.