Prompt gamma activation analysis using a focused thermal neutron guided beam at JAEA JRR-3M was applied to the determination of B in ceramic certified reference materials (BAM CRM S-003 Silicon Carbide Powder and NMIJ CRM 8004-a Silicon Nitride Powder). Cl and Si were used as internal standards to obtain linear calibration curves of B. The analytical result of B in BAM CRM S-003 was in good agreement with the certified value. The relative expanded measurement uncertainties (k = 2) were 4.8% for BAM CRM S-003 and 4.9% for NMIJ CRM 8004-a.