View More View Less
  • 1 Dipartimento di Chimica Fisica dell' Università' degli Studi di Pavia C.S.G.I. - Viale Taramelli 16 27100 Pavia Viale Taramelli 16 27100 Pavia
  • | 2 Unitá INFM Pavia Italy Pavia Italy
Restricted access

Cross Mark

Abstract  

The heat capacity calibration ‘constants’ of a commercial MTDSC system (TA 3100) were determined in a variety of experimental conditions. For a given modulation frequency, the calibration constants are the same within a few percents for different temperatures, and over a wide range of modulation amplitudes and scan rates. This variation decreases below 1% if hidden instrumental constraints are taken into account, which are related with the capability of the control system to achieve the desired temperature program. On the other hand, the calibration constant changes substantially with the period, and takes anomalously high values for the short modulation periods (20+40 s). Rules to optimize the accuracy of the system are given.

All Time Past Year Past 30 Days
Abstract Views 1 1 0
Full Text Views 1 1 0
PDF Downloads 0 0 0