Objective

We describe a method to stabilize CS lead position using stent implantation in a CS side branch to anchor the electrode to the wall of the vein, in cases of intraoperative or postoperative lead dislocation, unstable lead position and phrenic nerve stimulation (PNS).

Methods

403 patients were treated with stenting. After finding the desired lead position bare metal coronary stent was introduced via another guide wire, but in the same CS sheath. The stent was deposited 5–35 mm proximal to the tip of the electrode with a pressure of 6 to 14 atmospheres.

Results

Mechanical damage of the CS side branch or pericardial effusion was not observed. During follow-up (median 39, 23–48, max. 82 months) re-operation was necessary in only two patients because of high pacing threshold, while repositioning with ablation catheter was performed in 7 cases because of PNS. Impedance measurements did not suggest lead insulation failure. Transvenous extraction of stented CS leads was successful after 3, 18 and 49 months, while 4 leads were extracted easily during heart transplantation.

Conclusion

Stent implantation to stabilize CS lead position seems to be effective and safe for the prevention and treatment of CS lead dislocation in special cases.

  • 1. A.J. Moss W.J. Hall D.S. Cannom 2009 Cardiac-resynchronization therapy for the prevention of heart-failure events N Engl J Med 361 1329 1338.

    • Search Google Scholar
    • Export Citation
  • 2. D. Gras D. Bocker M. Lunati 2007 Implantation of cardiac resynchronization therapy systems in the CARE-HF trial: procedural success rate and safety Europace 9 516 522.

    • Search Google Scholar
    • Export Citation
  • 3. A.R. Leon W.T. Abraham A.B. Curtis 2005 Safety of transvenous cardiac resynchronization system implantation in patients with chronic heart failure: Combined results of over 2,000 patients from a multicenter study program J Am Coll Cardiol 46 2348 2356.

    • Search Google Scholar
    • Export Citation
  • 4. C.J. Borleffs R.J. van Bommel S.G. Molhoek J.G. de Leeuw M.J. Schalij L. van Erven 2009 Requirement for coronary sinus lead interventions and effectiveness of endovascular replacement during long-term follow-up after implantation of a resynchronization device Europace 11 607 611.

    • Search Google Scholar
    • Export Citation
  • 5. L. Gellér S. Szilágyi A. Róka B. Merkely 2005 Coronary sinus sidebranch stenting as a tool for coronary sinus lead dislocation in patients with heart failure and left bundle branch block Eur Heart J 26Suppl1 52/P486.

    • Search Google Scholar
    • Export Citation
  • 6. S. Szilagyi B. Merkely A. Roka 2007 Stabilization of the coronary sinus electrode position with coronary stent implantation to prevent and treat dislocation J Cardiovasc Electrophysiol 18 303 307.

    • Search Google Scholar
    • Export Citation
  • 7. O. Kowalski R. Lenarczyk J. Prokopczuk 2006 Effect of percutaneous interventions within the coronary sinus on the success rate of the implantations of resynchronization pacemakers Pace 29 1075 1080.

    • Search Google Scholar
    • Export Citation
  • 8. D.A. Cesario M. Shenoda R. Brar K. Shivkumar 2006 Left ventricular lead stabilization utilizing a coronary stent Pacing Clin Electrophysiol 29 427 428.

    • Search Google Scholar
    • Export Citation
  • 9. S. Szilágyi B. Merkely E. Zima L. Gellér 2011 Long-term experience with coronary sinus side branch stenting to stabilize left ventricular electrode position Heart Rhythm 8 845 850.

    • Search Google Scholar
    • Export Citation
  • 10. S. Szilágyi B. Merkely E. Zima L. Gellér 2008 Minimal invasive coronary sinus lead reposition technique for the treatment of phrenic nerve stimulation Europace 10 1157 1160.

    • Search Google Scholar
    • Export Citation
  • 11. E.W. Lau 2009 Achieving permanent left ventricular pacing — Options and choice PACE — Pacing and Clinical Electrophysiology 32 1466 1477.

    • Search Google Scholar
    • Export Citation
  • 12. M. Azizi M.A. Castel S. Behrens W. Rodiger H. Nagele 2006 Experience with coronary sinus lead implantations for cardiac resynchronization therapy in 244 patients Herzschrittmacherther Elektrophysiol 17 13 18.

    • Search Google Scholar
    • Export Citation
  • 13. M.R. Bristow L.A. Saxon J. Boehmer 2004 Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure N Engl J Med 350 2140 2150.

    • Search Google Scholar
    • Export Citation
  • 14. F.A. McAlister J.A. Ezekowitz N. Wiebe 2004 Systematic review: Cardiac resynchronization in patients with symptomatic heart failure Ann Intern Med 141 381 390.

    • Search Google Scholar
    • Export Citation
  • 15. J.G. Cleland J.C. Daubert E. Erdmann 2005 The effect of cardiac resynchronization on morbidity and mortality in heart failure N Engl J Med 352 1539 1549.

    • Search Google Scholar
    • Export Citation

Monthly Content Usage

Abstract Views Full Text Views PDF Downloads
May 2020 0 2 1
Jun 2020 0 4 0
Jul 2020 7 2 0
Aug 2020 9 0 0
Sep 2020 9 0 0
Oct 2020 5 0 0
Nov 2020 0 0 0