The warm-up effect is a well-known phenomenon, which occurs in all types of laser trackers. The series of experiments was performed to determine the influence of warm-up effect on measurement and a warm-up time of device – the time after the temperature inside the tracker is stable. In this paper, the tested tracker was Leica AT960-MR. Results showed that the warm-up effect could cause errors up to tenths of millimeters, and a warm-up time of instrument is around two hours, which is similar to the other researches.

  • [1]

    BS EN ISO 10360-10:2016, Geometrical product specifications (GPS) - Acceptance and re-verification tests for coordinate measuring systems (CMS) - Part 10: Laser trackers for measuring point-to-point distances, 2016.

    • Search Google Scholar
    • Export Citation
  • [2]

    Berényi A. Laser scanning in engineering survey–An application study, Pollack Periodica, Vol. 5, No. 2, 2010, pp. 3948.

  • [3]

    Honti R., Erdélyi J., Kopáčik A. Plane segmentation from point clouds, Pollack Periodica, Vol. 13, No. 2, 2018, pp. 159171.

  • [4]

    Dvořáček F. Laboratory testing of the Leica AT401 laser tracker, Acta Polytechnica, Vol. 56, No. 2, 2016, pp. 8898.

  • [5]

    Sugahara R., Masuzawa M., Ahsawa Y. Performance test of laser trackers of Faro, Proceedings of the 7th Annual Meeting of Particle Accelerator Society of Japan, Himeji, Japan, 4-6 August 2010, pp. 253256.

    • Search Google Scholar
    • Export Citation
  • [6]

    Ekinici O. 5 laser tracker warm-up tips for ideal measurement conditions, https://insights.faro.com/buildit-software/5-laser-tracker-warm-up-tips-for-ideal-measurement-conditions,(last visited 12 December 2019).

    • Search Google Scholar
    • Export Citation
  • [7]

    Accurate measurement best practices for the laser trackers, https://knowledge.faro.com/Hardware/Laser_Tracker/Tracker/Accurate_Measurement_Best_Practices_for_the_Laser_Tracker, (last visited 12 December 2019).

  • [8]

    Automated precision, Inc. Tracker 3, Laser tracking system, user’s manual, v. 1.0, 2005.

  • [9]

    Gassner G., Ruland R. Instrument tests with the new leica AT401, International Workshop on Accelerator Alignment, Stanford, CA, USA, 5 January 2011, pp. 15.

    • Search Google Scholar
    • Export Citation
  • [10]

    Renishaw, interferometry explained, https://www.renishaw.com/en/interferometry-explained--7854, (last visited 12 December 2019).

  • [11]

    Muralikrishnan B., Phillips S., Sawyer D. Laser trackers for large-scale dimensional metrology: A review, Precision Engineering, Vol. 44, 2016, pp. 1328,

    • Crossref
    • Search Google Scholar
    • Export Citation
  • [12]

    Harding K. Handbook of optical dimensional metrology, CRC Press, Florida, 2008.

  • [13]

    Leica Geosystems, Leica absolute tracker AT401, White Paper, Switzerland, 2010.

  • [14]

    ASME B89.4.19:2006, Performance evaluation of laser-based spherical coordinate measurement systems, 2006.