Abstract
Olmutinib (Olita™) is an oral third-generation epidermal growth factor receptor tyrosine kinase inhibitor (EGFR-TKI) which is used to treat non-small cell lung cancer (NSCLC). A simple, rapid, and sensitive method based on ultra-performance liquid chromatography tandem mass spectrometry (UHPLC–MS/MS) has been developed for the determination of olmutinib. Sample preparation was performed following simple one-step protein precipitation with acetonitrile. Olmutinib and internal standard (dasatinib) were separated on an Eclipse Plus C18 RRHD (2.1 × 50 mm, 1.8 μm) column. The mobile phase consisted of acetonitrile–0.1% formic acid in water with gradient elution. A total run time of 1.7 min was achieved. Detection was performed on a positive-ion electrospray ionization mass spectrometer in multiple reaction monitoring (MRM) mode, using transitions of m/z 487.2 → 402.1 for olmutinib and m/z 488.2 → 401 for dasatinib (IS), respectively. The calibration curve (R2 = 0.999) was linear over the range of 1–500 ng/mL. The recovery of olmutinib ranged from 85.8% to 95.5%. This method can be applied to pharmacokinetic studies of olmutinib.
Introduction
Lung cancer is the leading cause of cancer death worldwide. Non-small cell lung cancer (NSCLC) is the most common type, accounting for more than 80% of all lung cancers [1, 2]. NSCLC includes squamous cell carcinoma, adenocarcinoma, and large cell carcinoma. Compared with small cell lung cancer, NSCLC cells proliferate more slowly and metastasize later. Despite novel diagnostic and therapeutic advances, the 5-year survival rate for NSCLC remains approximately 15% [3].
Approximately 19% to 51.4% of NSCLC cases demonstrate overexpression of epidermal growth factor receptor (EGFR) [4, 5]. Some studies suggest that EGFR-activated cell signaling plays a significant role in lung tumorigenesis and tumor progression [6, 7]. Therefore, inhibition of EGFR signaling has been suggested as an approach to NSCLC treatment [8–10]. Several EGFR tyrosine kinase inhibitors (EGFR-TKIs) have been developed. First-generation EGFR-TKIs such as gefitinib and erlotinib give good primary responses against activating EGFR mutations by binding to the tyrosine kinase (TK) domain of EGFR. Unfortunately, many patients acquire resistance during the first year of treatment [11, 12]. Second-generation EGFR-TKIs such as afatinib and dacomitinib bind irreversibly to the TK domain of the EGFR and other receptors of the ERbB family [12, 13]. These newer agents also do not prevent development of tumor resistance. This leads to the development of third-generation TKIs that target the EGFR T790M mutation, an important resistance mechanism against EGFR-TKIs [12, 13].
Olmutinib (Olita™) is an oral third-generation EGFR mutant-selective TKI developed by Hanmi Pharmaceutical Co. Ltd. It was approved in South Korea on May 13, 2016 (Figure 1) [14]. Oral olmutinib is absorbed well. In EGFR mutation-positive NSCLC patients, median time to maximum serum concentration is 2.5–5.9 h. Mean half-life is 8.1–18.3 h with a dosage range of 75 to 1200 mg/day. In phase I/II trials, adverse events (AEs) caused by olmutinib included diarrhea and pruritis, but there were no reports of hyperglycemia, QT prolongation, or treatment-related deaths [14].
Based on the above background, olmutinib will be a hot spot in the field of NSCLC in the future and, as a new drug, more and more experiments about olmutinib will be done. However, a method for determination and pharmacokinetic investigation of olmutinib has not yet been reported. In this article, we describe a quantitative analysis of olmutinib using UHPLC–MS/MS.
Experimental
Compounds
Olmutinib (>99% pure) was obtained from Perfemiker. The internal standard (IS) dasatinib was acquired from Toronto Research Chemicals (TRC). Methanol (liquid chromatography–mass spectrometry [LC–MS] grade) and acetonitrile (LC–MS grade) were supplied by Merck Company (Darmstadt, Germany). About 0.1% formic acid (99.9% purity) was obtained from J & K Scientific Ltd. (Shanghai, China). Ultrapure water was prepared by a Milli-Q system (Millipore, Bedford, MA).
Chromatography and MS/MS Method
Chromatographic analysis was performed on an Agilent 1290 system (Agilent Technologies, USA). Partial loop injections of 2 μL were made on an Eclipse Plus C18 RRHD (2.1 × 50 mm, 1.8 μm) column. The compounds were eluted by 0.1% formic acid in water (A) and acetonitrile (B). The 1.7 min gradient was run as follows: 27–95% B (0–0.6 min), maintained at 95% B (0.6–1.6 min), and 95–27% B (1.6–1.7 min). The total run time (3.3 min) included stop time (1.7 min) and post time (1.6 min). The flow rate was 0.4 mL/min.
The sample was tested on an Agilent G6420A Triple quadrupole LC–MS (manufactured in Singapore using domestic and foreign components) equipped with an electrospray ionization (ESI) source. The fragmentor voltage was 180 V for olmutinib and 214 V for dasatinib. Olmutinib was monitored at m/z 487.2 → 402.1 at 38 V collision energy, and dasatinib (IS), at m/z 488.2 → 401 at 33 V collision energy. The gas temperature was set to 350 °C, and the flow rate was 10 L/min. The cell accelerator voltage for olmutinib was 3.0 kV. Dasatinib voltage was 5.0 kV.
Mass Hunter work station data acquisition software was used for data collection, and Qualitative Analysis software (version B.07.00) was used to process the data.
Preparation of Standard Solutions
A stock solution of olmutinib was prepared in methanol, which was then diluted with methanol to various concentrations as the working solution. Calibration standards and quality control (QC) samples were prepared by adding the above working solutions to blank rat plasma. Final calibration standards of olmutinib were prepared at 1, 2, 5, 10, 20, 50, 100, 200, and 500 ng/mL. QC samples were set at 3 ng/mL, 40 ng/mL, and 400 ng/mL.
Sample Preparation
Ninety microliters of rat plasma and 10 μL various concentrations working solutions of olmutinib were mixed in a 1.5-mL tube. Then, 20 μL dasatinib (IS, 10 μg/mL) and 200 μL acetonitrile were added. After mixing well and centrifuging, 150 μL supernatant was collected and diluted by pure water for testing.
Pharmacokinetic Study
Six male Sprague-Dawley rats (240–310 g) were provided by the Laboratory Animal Center of Wenzhou Medical University (Wen-Zhou, China) to perform the pharmacokinetic study of olmutinib. All rats were fasted overnight. Olmutinib was dissolved in dimethyl sulfoxide (DMSO) (modicum, since DMSO is toxic) and edible oil. Following an oral dose of 50 mg/kg, 300 μL of blood samples was drawn from rat tail vein. Samples were obtained at 10 min, 30 min, 1 h, 2 h, 4 h, 6 h, 8 h, 10 h, 12 h, 24 h, and 36 h.
Method Validation
Validation of the method was carried out according to a laboratory scheme published by the Food and Drug Administration (FDA) and European Medicines Agency (EMA).
A nine-point calibration curve (1–500 ng/mL) was analyzed by means of the least squares linear regression (1/×2 weighted), using the peak area ratio against concentrations. Coefficient (r2) values was 0.999, which was better than 0.985. The lower limit of quantification (LLOQ) was determined as the lowest concentration on the calibration curves.
Intra-day and inter-day accuracy and precision were determined by analysis of six replicates of QC samples over 3 days. Relative standard deviation (RSD %) was calculated for precision, while relative error (RE %) was determined for accuracy. Both RSD % and RE % are expected to be less than ±15%.
Selectivity of this method was analyzed by comparing the results of six batches of blank plasma samples, rat plasma samples, and blank sample containing olmutinib and dasatinib (IS). The purpose of the analysis was to test for the presence of endogenous interference introduced by blank plasma.
To evaluate the recovery of olmutinib, two types of peak areas were examined. One was from six QC samples to which olmutinib was added prior to the precipitation reaction of blank plasma. The second was obtained from samples to which olmutinib was added following protein precipitation. The matrix effect was analyzed according to the ratio of the peak areas of the analytes containing the same amount of blank rat plasma dissolved in standard solution.
The stability of olmutinib in plasma samples was determined under different storage conditions as follows: maintained room temperature for 4 h and at 4 °C for 24 h, cryopreserved at −20 °C for 15 days, and processed through three freeze–thaw cycles (thawing at room temperature and refreezing at −20 °C).
Results and Discussion
Method Development
A reliable mass spectrometry method was established based on various previous trials. MS responses for olmutinib and dasatinib (IS) were better in positive ionization mode versus negative ionization mode. To optimize the mass spectrometry method, collision energy (CE) was adjusted from 5 to 45, and fragmentor voltage was changed from 50 V to 220 V. As a result, olmutinib was monitored at m/z 487.2 → 402.1 at 38 V collision energy, and dasatinib (IS) was at m/z 488.2 → 401 at 33 V collision energy. More details of spectrometric parameters are displayed in Table 1 and Figure 2.
MS parameters for olmutinib and dasatinib
Compound name | Precursor ion (m/z) | Product ion (m/z) | Collision energy | Fragmentor |
---|---|---|---|---|
Olmutinib | 487.2 | 402.1 | 38 | 180 |
Dasatinib | 488.2 | 401 | 33 | 214 |
Chromatographic conditions included constitution of mobile phase and column. Liquid phase schemes included water, acetonitrile, and water with 0.1% formic acid. The result was that acetonitrile and water with 0.1% formic gave good peak shape without distinct peak broadening or tailing. Gradient elution, as opposed to isocratic elution, was used to obtain shorter retention times. Flow rate of liquid phase was tried between 0.2 and 0.4 mL/min. Column temperature changed from 25 to 45 °C.
Method Validation
The calibration curve of olmutinib (n = 9) was obtained from the concentration range of 1–500 ng/mL, with a coefficient of determination R2 = 0.999, demonstrating good linearity. The LLOQ was determined as the lowest concentration on the calibration curves.
Intra-day and inter-day RSD (representing precision) as well as intra-day and inter-day RE (representing accuracy) were within the ±15% range as required. The extraction recoveries were good, and the matrix effect could be ignored. All the details were showed in Table 2.
Precision and accuracy for olmutinib of QC sample in rat plasma (n = 6)
Analytes | Level (ng/mL) | Intra-day precision | Inter-day precision | ||
---|---|---|---|---|---|
RE (%) | RSD (%) | RE (%) | RSD (%) | ||
Olmutinib | 3 ng/mL | 2.0 | 10.0 | −1.1 | 2.2 |
40 ng/mL | −9.8 | 1.8 | −11.9 | 5.8 | |
400 ng/mL | 12.0 | 3.3 | 10.5 | 1.4 |
To evaluate selectivity of the assay, blank rat plasma samples, blank rat plasma mixed with olmutinib and dasatinib (IS), and a rat plasma sample were analyzed. The chromatograms obtained from these samples are shown in Figure 3. No obvious interferences were found at the retention time of olmutinib (1.4 min) and internal standard (0.8 min).
The values of the extraction recoveries and the matrix effects of olmutinib were showed in Table 3. The stability of olmutinib was tested under different conditions. QC samples were kept at room temperature for 4 h, at 4 °C for 24 h, and processed in three freeze–thaw cycles to examine short-term stability. Cryopreservation at −20 °C for 15 days was examined to test long-term stability. The results were displayed in Table 4, demonstrating that no obvious change of olmutinib appeared in rat plasma in all storage conditions.
Recovery and matrix effect for olmutinib of QC sample in rat plasma (n = 6)
Analytes | Level (%) | Recovery (%) | Matrix effect (%) |
---|---|---|---|
Olmutinib | 3 ng/mL | 85.8 | 102.3 |
40 ng/mL | 95.8 | 96.9 | |
400 ng/mL | 95.5 | 96.9 |
Summary of stability of olmutinib in rat plasma under different storage conditions (n = 6)
Analytes | Level (%) | Room temperature | 4 °C | Three freeze–thaw | −20 °C | ||||
---|---|---|---|---|---|---|---|---|---|
RE (%) | RSD (%) | RE (%) | RSD (%) | RE (%) | RSD (%) | RE (%) | RSD (%) | ||
Olmutinib | 3 ng/mL | 6.3 | 11.0 | 5.4 | 12.9 | 7.5 | 4.5 | 7.3 | 10.8 |
40 ng/mL | −13.4 | 7.8 | −8.3 | 2.1 | −4.5 | 4.5 | 5.0 | 5.9 | |
400 ng/mL | 11.8 | 3.0 | 8.9 | 1.6 | 1.0 | 2.6 | 8.9 | 1.6 |
Method Application
The developed UHPLC–MS/MS method was successfully applied to the plasma sample analysis in a pharmacokinetic study of olmutinib. Six SD rats were employed to obtain a concentration–time curve (Figure 4) following oral administration of olmutinib (50 mg/kg). Pharmacokinetic parameters were analyzed by DAS software (version 3.0) using non-compartmental mode. Complete data are displayed in Table 5.
The pharmacokinetic parameters of olmutinib in rat plasma after oral administration
Parameters | Unit | Mean | SD |
---|---|---|---|
po 50 mg/kg | |||
AUC(0–t) | μg/L*h | 4562.5 | 975.5 |
AUC(0–∞) | μg/L*h | 4584.2 | 971.0 |
t 1/2 | h | 3.6 | 1.4 |
T max | h | 7.7 | 0.82 |
V | L /kg | 60.3 | 32.5 |
CL | L/h/kg | 11.9 | 2.4 |
C max | μg/L | 597.4 | 68.5 |
Conclusions
A UHPLC–MS/MS method with good selectivity and sensitivity has been successfully established and validated. This method is simple, rapid, and sensitive, with good linearity, precision, accuracy, recovery, and stability. The method was successfully employed in a pharmacokinetic study in rats. Furthermore, this method can be used as a reference for the therapeutic drug monitoring. To our knowledge, this is the first report of a validated assay for olmutinib.
References
- 1.↑
Torre, L. A.; Bray, F.; Siegel, R. L.; Ferlay, J.; Lortet-Tieulent, J.; Jemal, A. Ca-Cancer J. Clin. 2015, 65, 87–108.
- 4.↑
Weiss, J. M.; Villaruz, L. C.; Socinski, M. A.; Ivanova, A.; Grilley-Olson, J.; Dhruva, N.; Stinchcombe, T. E. Lung Cancer 2014, 86, 288–290.
- 5.↑
Shi, Y.; Au, J. S.; Thongprasert, S.; Srinivasan, S.; Tsai, C. M.; Khoa, M. T.; Heeroma, K.; Itoh, Y.; Cornelio, G.; Yang, P. C. J. Thorac. Oncol. 2014, 9, 154–162.
- 7.↑
Chou, Y. T.; Lin, H. H.; Lien, Y. C.; Wang, Y. H.; Hong, C. F.; Kao, Y. R.; Lin, S. C.; Chang, Y. C.; Lin, S. Y.; Chen, S. J.; Chen, H. C.; Yeh, S. D.; Wu, C. W. Cancer Res. 2010, 70, 8822–8831.
- 9.
Glover, K. Y.; Perez-Soler, R.; Papadimitradopoulou, V. A. Semin. Oncol. 2004, 31, 83–92.
- 10.↑
Haber, D. A.; Bell, D. W.; Sordella, R.; Kwak, E. L.; Godin-Heymann, N.; Sharma, S. V.; Lynch, T. J.; Settleman, J. Cold Spring Harbor Symp. Quant. Biol. 2005, 70, 419–426.