Metaldehyde is a molluscicide allowed for use in the control of slugs and snails in agriculture and horticulture in many countries. A simple, fast, and precise gas chromatography method was developed and single-laboratory validated for determination of metaldehyde in different formulations of plant protection products. The proposed method involves extraction of active substance from samples by sonication with acetone and analysis using gas chromatography–flame ionization detection (GC–FID). The suggested analytical procedure is accurate, precise, and repeatable. Moreover, it is environmentally friendly and useful for laboratories as it uses a no time- and no solvent-consuming reference chromatography technique for quality control of commercially available pesticide formulations. Advantages of the proposed method are consistent with the ideas of sustainable development, which are in accordance with the principles of Green Analytical Chemistry. Analysis of real samples of commercial pesticide formulations confirmed that the proposed method is fit for its purpose.

Abstract

Metaldehyde is a molluscicide allowed for use in the control of slugs and snails in agriculture and horticulture in many countries. A simple, fast, and precise gas chromatography method was developed and single-laboratory validated for determination of metaldehyde in different formulations of plant protection products. The proposed method involves extraction of active substance from samples by sonication with acetone and analysis using gas chromatography–flame ionization detection (GC–FID). The suggested analytical procedure is accurate, precise, and repeatable. Moreover, it is environmentally friendly and useful for laboratories as it uses a no time- and no solvent-consuming reference chromatography technique for quality control of commercially available pesticide formulations. Advantages of the proposed method are consistent with the ideas of sustainable development, which are in accordance with the principles of Green Analytical Chemistry. Analysis of real samples of commercial pesticide formulations confirmed that the proposed method is fit for its purpose.

Introduction

Metaldehyde (2,4,6,8-tetramethyl-1,3,5,7-tetraoxacyclo-octane) is a tetramer of acetaldehyde. Information about the slug-killing properties of metaldehyde was first published in 1936 [1]. It is a molluscicide with contact and stomach action—poisoned slugs secrete large quantities of slime, dehydrate, and die—which has been used in the control of slugs and snails in agriculture and horticulture. Plant protection products with metaldehyde are produced in the form of granular bait (GB) and granule (GR), as well as a ready-to-use bait (RB) [1].

Data from January 2018 indicate that metaldehyde is approved for use in twenty five European Union (EU) countries [2]. Analysis of data regarding plant protection products approved for use in Poland in the most recent years indicates a significant increase in the number of products containing metaldehyde as an active substance, which are available on the Polish market. Figure 1 shows the number of metaldehyde preparations placed on the market in Poland in 2004–2017 (as of January each year) [3]. There was a sharp increase in the number of issued permits in 2015–2017, whereas from 2004 to 2014, the number of preparations available on the Polish market remained stable. In Poland, according to the registry of January 2018, pursuant to a permit issued by the Minister of Agriculture and Rural Development, there were 29 preparations approved for use, including 23 GB preparations, 4 RB preparations, and 2 GR products. These products contain 25–50 g/kg of metaldehyde [4]. The number of products is about half way in comparison with countries like Germany or Great Britain [5, 6].

Figure 1.
Figure 1.

The number of metaldehyde preparations registered for use in Poland in 2004–2017

Citation: Acta Chromatographica Acta Chromatographica 31, 4; 10.1556/1326.2018.00510

Since metaldehyde formulations have been used for a number of years, a range of methods were developed and published for the determination of metaldehyde residues in different matrices including water, human serum, fruits, and vegetables. These include simple systems like gas chromatography with flame ionization detector (GC–FID) employed for determination of metaldehyde in animal materials or nitrogen-phosphorus detector (NPD) used upon prior metaldehyde derivatization in citrus fruit and tobacco plants [79]. Metaldehyde residues were also determined using more complex systems like gas chromatography coupled with mass spectrometry (GC–MS) (determination of metaldehyde in water and riverbed deposits in the case of metaldehyde poisonings and residue analyses in foods) [1013], liquid chromatography coupled with tandem mass spectrometry (LC–MS/MS) (metaldehyde determination in vegetables, soil, and water) [1419], and headspace gas–chromatography coupled with mass spectrometry (HS–GC–MS) (metaldehyde determination in human plasma) [20], as well as gas chromatography with a less popular rubidium sulfate alkali-flame ionization detector (AFD) [21].

The authors of this article are not aware of any recent publications concerning metaldehyde determination in formulations of plant protection products, which are currently available on the market. The subject matter literature references a method of metaldehyde determination in plant protection products using GC–MS with m-xylene as an internal standard and trichloromethane as a solvent (the authors were not able to access the full description of the above-mentioned method) [1]. This method, however, does not correlate in any way with the ideas of sustainable development, currently taking hold in a growing number of laboratories, in the form of the principles of Green Analytical Chemistry (GAC). According to the principles, one should (1) prefer direct techniques so as to eliminate the step of sample preparation for analysis, (2) reduce the number and the volume of collected samples, (3) perform in situ measurements if possible, (4) integrate processes and analytical operations to reduce reagents and energy consumption, (5) employ automated and miniaturized methods, (6) resign from derivatization, (7) aim at reduction of generated analytical wastes and their proper neutralization, (8) select techniques enabling multi-analyte determination or multi-parameter measurement in a single run, (9) minimize energy consumption, (10) avoid using toxic reagents, and (11) increase the work safety of analytical chemists [2225]. It is worth underlining that the European Union law according to Article 68 of the Regulation (EC) No 1107/2009 of the European Parliament and of the Council of 21 October 2009, concerning the placing of plant protection products on the market and repealing Council Directives 79/117/EEC and 91/414/EEC, mandates the member states to perform official control of pesticides which are being introduced on the market [26]. A key component of the control is the quality testing of samples taken among commercially available products, verifying that the products are of adequate quality, i.e., they comply with the requirements and criteria set during the registration process. The control measure is designed to prevent the use of non-compliant pesticides, which could have an adverse or harmful effect on crops, animals, the environment, and humans [26].

The aim of the study was to develop a simple, fast, inexpensive, and GAC-compliant method for the determination of metaldehyde, which could be easily adopted in routine activities of a quality control laboratory.

Experimental

2.1. Reagents and Materials

The analytical standard of metaldehyde (99.0%) was obtained from Dr. Ehrenstorfer (currently LGC) (Augsburg, Germany). Acetone (for residue pesticide analysis) was purchased from Avantor Performance Materials Poland S.A. (Gliwice, Poland). Commercial formulations containing metaldehyde in the form of GB, GR, and RB were acquired from the Polish market. The plant protection products with metaldehyde available in Poland at the time, when this work was carried out, contained 25–50 g/kg of active substance.

2.2. Preparation of Reference Solutions

Solutions of analytical standard of metaldehyde were prepared at the concentrations of 1 mg/mL by weighting an appropriate amount of compound and diluting it with acetone.

2.3. Sample Preparation

For the determination of the active substance, about 10 g of a formulation was weighted on a precision balance and ground using the pulse function of an electric laboratory grinder. A part of the pre-ground sample was then transferred into a mortar and further processed into fine powder. Then, an appropriate amount of a commercial formulation was weighted to obtain the concentration of approximately 1 mg/mL of the active substance in the solution. Three portions of each sample were weighted and diluted with acetone to the final volume, then sonicated for 10 min and allowed to cool to room temperature. Prior to gas chromatography analysis, cooled samples were filtered into chromatographic vials with the use of polytetrafluoroethylene (PTFE) filters (0.45 μm). Analysis of each sample was carried out in triplicate.

2.4. Instrumentation and Conditions

Separation, identification, and quantification were carried out using an Agilent 7890 series gas chromatograph, equipped with flame ionization detector (FID) (Agilent Technologies, Palo Alto, CA, USA). Analyte was separated with an HP-5 capillary column (30 m × 0.25 mm × 0.25 μm) (Agilent Technologies, Folsom, CA, USA). Extracts (1 μL) were injected in the split mode (split ratio of 1:50) using a standard split/splitless injector kept at the temperature of 280 °C. High purity helium (99.999%, Messer Group GmbH, Bad Soden, Germany) was used as carrier gas with a constant flow of 1.76 mL/min. The column oven temperature was held at a constant temperature of 100 °C. The total run time was 10 min. The detector temperature was kept at 300 °C. Instrument control and data processing was carried out using an Agilent GC ChemStation B.04.03 (Agilent Technologies, Palo Alto, CA, USA).

Results and Discussion

Method validation was realized according to SANCO/3030/99 rev. 4 and CIPAC guidelines on the basis of the following parameters: specificity, linearity, precision, and accuracy. Generally blank formulations of plant protection products are not available; therefore, a standard addition method was used for estimation of accuracy. In accordance with the above-mentioned guidelines, establishing limits of quantification (LOQ) is not required for an active substance, and hence it was not performed [2729].

Specificity is the ability of a method to distinguish between the analyte being measured and other substances, based upon sufficient characteristics of the analyte as to make the results completely specific to the analyte, irrespective of the characteristics of other materials present [27]. Since blank pesticide formulations were not available, the lack of interference was demonstrated in the analysis of formulations samples, solvent, and standard solutions. Figure 2 shows a typical result of separation of detected compounds.

Figure 2.
Figure 2.

An example of a GC–FID chromatogram for a sample of a plant protection product containing metaldehyde (tR = 3.403 min)

Citation: Acta Chromatographica Acta Chromatographica 31, 4; 10.1556/1326.2018.00510

Linearity is the ability of a detection system, within a defined range, to produce an acceptable linear correlation between the test results and the concentration of analyte in the sample. The linearity of response was assessed, considering the range of nominal concentration of the target analyte in the analyzed matrices ±30% (according to the requirements, the minimum range should be ±20%). Five working standard solutions of metaldehyde were prepared at concentrations of 0.70, 0.85, 1.00, 1.15, and 1.30 mg/mL. Data was collected from six injections at each calibration level. Linear regression performed with the use of least squares method demonstrated satisfactory linear relationship of the analyte response with correlation coefficient ≥0.99. The obtained linearity parameters are summarized in Table 1.

Table 1.

Linearity parameters obtained for GC-FID analysis of metaldehyde in plant protection products formulations

SubstanceCalibration levelsSlope a ± SDaIntercept b ± SDbCorrelation coefficientSy/xaConcentration range [mg/mL]
Metaldehyde5150.93 ± 6.37−4.83 ± 6.510.9973.020.7–1.3

Residual standard deviation.

The precision of the method was determined by analysis of six different sample solutions prepared from the same batch of a pesticide formulation. Suitability criteria of precision were based on the modified Horwitz equation [27]. The obtained experimental RSDr value was compared with the RSDr value calculated from the modified Horwitz equation (Eq. 1).
RSDr%=RSDR%×0.67
where
RSDR%=210.5logC
where C is the concentration of an analyte in a sample expressed as a decimal mass fraction, RSDr is the repeatability relative standard deviation, and RSDR is the inter-laboratory relative standard deviation.

Precision was considered acceptable if the experimental RSDr value was lower than the calculated RSDr value. Precision assessed for metaldehyde determinations in plant protection products is given in Table 2. The obtained results demonstrate adequacy of the proposed method for the intended application.

Table 2.

Precision for tested plant protection products

Plant protection productDeclared concentration of metaldehyde [%]% RSDr teoretical% RSDr experimental
Product A formulation GB5.02.101.15–2.07
Product B formulation GR3.02.271.35–1.37
Product C formulation GB3.00.81
Product D formulation GB3.01.34–1.68
Product E formulation GB3.00.55–1.14
Product F formulation GB3.00.84–1.04
Product G formulation RB2.52.331.16
Product H formulation GB3.02.272.05–2.11

Accuracy is the degree to which the determined concentration/content of an analyte in a sample corresponds to the true value [27]. In this study, blank (i.e., analyte-free) formulations were not available, and the accuracy of measurements was determined by recovery studies applying the standard addition method. Three portions of each pesticide formulation in acetone containing 1 mg of metaldehyde per 1 mL of solution were prepared, separately for GB, GR, and RB formulations. Then, 0.0, 1.5, and 3.0 mL of the standard solution with appropriate concentration of metaldehyde were added, and finally, the mixture was diluted to 10 mL with acetone. Recoveries were determined in three replicates at two spiking levels. The spiking level depended on the amount of the active substance present in the formulation. According to the SANCO guidelines, a spiking range of 1–10% should give recovery of 98–102%. All obtained recovery values fell within the respective acceptance ranges (Table 3).

Table 3.

Accuracy expressed as the percent recovery of a spike

Plant protection productDeclared concentration of metaldehyde [%]Accepted recovery [%]Recovery [%]
Product A formulation GB5.097–103100.65–102.28
Product B formulation GR3.0100.27–101.98
Product C formulation GB3.097.18–98.61
Product D formulation GB3.0100.02–101.24
Product E formulation GB3.099.47–101.68
Product F formulation GB3.097.40–99.64
Product G formulation RB2.597.46–98.76
Product H formulation GB3.097.41–100.11

Data obtained for measurements of the spiked samples were also used to determine the linearity of response. The linearity fit was good and equal to minimum r = 0.9981 for GB formulation, r = 0.9929 for GR, and r = 0.9934 for RB formulation, which indicated that the instrument response to the analyte in the spiked samples resulted solely from the change in the analyte concentration. This also provided a confirmation of the specificity of the method, as the contribution of matrix components to the instrument response was negligible.

In conclusion, validation of the developed method showed highly satisfactory results for all the required parameters of analytical performance.

To demonstrate applicability of the developed and validated method, real samples of commercial pesticide formulations containing 25 g/kg to 50 g/kg of metaldehyde were analyzed to determine the concentration of the active substance. An examplary GC–FID chromatogram is shown in Figure 2. Eighteen formulations containing metaldehyde available on the Polish market were analyzed; the obtained results are shown in Table 4.

Table 4.

Concentration of metaldehyde in difference plant protection products

Plant protection productNumber of analyzed pesticidesDeclared concentration of metaldehyde [%]Concentration of metaldehyde in analysed samples [%]a ± SDAccepted range of metaldehyde [%]
Product A formulation GB45.04.84 ± 0.06[4.8–5.2]
4.93 ± 0.08
5.07 ± 0.08
5.16 ± 0.06
Product B formulation GR33.03.01 ± 0.03[2.3–3.3]
3.05 ± 0.03
3.08 ± 0.04
Product C formulation GB13.02.71 ± 0.02[2.3–3.3]
Product D formulation GB23.02.77 ± 0.05[2.3–3.3]
3.04 ± 0.07
Product E formulation GB33.02.73 ± 0.02[2.3–3.3]
2.81 ± 0.02
2.88 ± 0.04
Product F formulation GB23.02.73 ± 0.04[2.3–3.3]
3.08 ± 0.04
Product G formulation RB12.52.54 ± 0.02[2.25–2.75]
Product H formulation GB23.02.82 ± 0.03[2.3–3.3]
3.13 ± 0.02

Each reported value is the mean ± SD (n = 6)

The tested formulations were produced by different manufacturers and represent eight products marked by letters A–H. The results of determination indicated that the active substance was within the acceptable range for all the samples.

Conclusions

The proposed GC–FID method for quality control of plant protection products with metaldehyde as an active substance was developed and validated according to the EU reference document recommendations [27]. The proposed approach can be easily implemented in control laboratories of member states. Also, the utility of the developed procedure can be verified by laboratories of third countries. Such compliance combined with the fact that the method provides for a fast, single analytical course results related to both the quality and quantity of an active substance makes the proposed method suitable for control testing of metaldehyde pesticides by laboratories working according to EU standards. It should also be underlined that the outline of a method for the determination of metaldehyde mentioned in the Pesticide Manual recommends using expensive and complicated analytical techniques (i.e., GC–MS) and hazardous chemicals (i.e., m-xylene and trichloromethane) [1]. The developed method could be relatively easily adopted even in not-very-well-equipment laboratories because it uses GC–FID, which is an easy, fast, and inexpensive technique. It is also worth mentioning that the proposed method for the determination of metaldehyde utilizes small amounts of low-toxicity reagents, i.e., acetone, involves an uncomplicated sample preparation step and uses a relatively simple, inexpensive, and commonly available analytical technique. Therefore, it is in line with the idea of sustainable development propagated, among others, by the so-called Green Analytical Chemistry principles. The proposed approach allowed the determination of metaldehyde in a complex mixture without interference from other components including degradation products, impurities, and formulation additives.

References

The author instruction is available in PDF.
Please, download the file from HERE.
 
The Open Access statement together with the description of the Copyright and License Policy are available in PDF.
Please, download the file from HERE.
 

 

Senior editors

Editor(s)-in-Chief: Kowalska, Teresa

Editor(s)-in-Chief: Sajewicz, Mieczyslaw

Editors(s) Danica Agbaba (University of Belgrade, Belgrade, Serbia);
Ivana Stanimirova-Daszykowska (University of Silesia, Katowice, Poland),
Monika Waksmundzka-Hajnos (Medical University of Lublin, Lublin, Poland)

Editorial Board

R. Bhushan (The Indian Institute of Technology, Roorkee, India)
J. Bojarski (Jagiellonian University, Kraków, Poland)
B. Chankvetadze (State University of Tbilisi, Tbilisi, Georgia)
M. Daszykowski (University of Silesia, Katowice, Poland)
T.H. Dzido (Medical University of Lublin, Lublin, Poland)
A. Felinger (University of Pécs, Pécs, Hungary)
K. Glowniak (Medical University of Lublin, Lublin, Poland)
B. Glód (Siedlce University of Natural Sciences and Humanities, Siedlce, Poland)
K. Kaczmarski (Rzeszow University of Technology, Rzeszów, Poland)
H. Kalász (Semmelweis University, Budapest, Hungary)
I. Klebovich (Semmelweis University, Budapest, Hungary)
A. Koch (Private Pharmacy, Hamburg, Germany)
Ł. Komsta (Medical University of Lublin, Lublin, Poland)
P. Kus (Univerity of Silesia, Katowice, Poland)
D. Mangelings (Free University of Brussels, Brussels, Belgium)
E. Mincsovics (Corvinus University of Budapest, Budapest, Hungary)
G. Morlock (Giessen University, Giessen, Germany)
J. Sherma (Lafayette College, Easton, PA, USA)
R. Skibiński (Medical University of Lublin, Lublin, Poland)
B. Spangenberg (Offenburg University of Applied Sciences, Germany)
T. Tuzimski (Medical University of Lublin, Lublin, Poland)
Y. Vander Heyden (Free University of Brussels, Brussels, Belgium)
A. Voelkel (Poznań University of Technology, Poznań, Poland)
B. Walczak (University of Silesia, Katowice, Poland)
W. Wasiak (Adam Mickiewicz University, Poznań, Poland)

KOWALSKA, TERESA
E-mail: kowalska@us.edu.pl

SAJEWICZ, MIECZYSLAW
E-mail:msajewic@us.edu.pl

Indexing and Abstracting Services:

  • Science Citation Index
  • Sci Search
  • Research Alert
  • Chemistry Citation Index and Current Content/Physical
  • Chemical and Earth Sciences
  • SCOPUS
  • GoogleScholar
  • Index Copernicus
  • CABI
2020
 
Total Cites
650
WoS
Journal
Impact Factor
1,639
Rank by
Chemistry, Analytical 71/83 (Q4)
Impact Factor
 
Impact Factor
1,412
without
Journal Self Cites
5 Year
1,301
Impact Factor
Journal
0,34
Citation Indicator
 
Rank by Journal
Chemistry, Analytical 75/93 (Q4)
Citation Indicator
 
Citable
45
Items
Total
43
Articles
Total
2
Reviews
Scimago
28
H-index
Scimago
0,316
Journal Rank
Scimago
Chemistry (miscellaneous) Q3
Quartile Score
 
Scopus
393/181=2,2
Scite Score
 
Scopus
General Chemistry 215/398 (Q3)
Scite Score Rank
 
Scopus
0,560
SNIP
 
Days from
58
sumbission
 
to acceptance
 
Days from
68
acceptance
 
to publication
 
Acceptance
51%
Rate

2019  
Total Cites
WoS
495
Impact Factor 1,418
Impact Factor
without
Journal Self Cites
1,374
5 Year
Impact Factor
0,936
Immediacy
Index
0,460
Citable
Items
50
Total
Articles
50
Total
Reviews
0
Cited
Half-Life
6,2
Citing
Half-Life
8,3
Eigenfactor
Score
0,00048
Article Influence
Score
0,164
% Articles
in
Citable Items
100,00
Normalized
Eigenfactor
0,05895
Average
IF
Percentile
20,349
Scimago
H-index
26
Scimago
Journal Rank
0,255
Scopus
Scite Score
226/167=1,4
Scopus
Scite Score Rank
Chemistry (miscellaneous) 240/398 (Q3)
Scopus
SNIP
0,494
Acceptance
Rate
41%

 

Acta Chromatographica
Publication Model Online only
Gold Open Access
Submission Fee none
Article Processing Charge 400 EUR/article
Regional discounts on country of the funding agency World Bank Lower-middle-income economies: 50%
World Bank Low-income economies: 100%
Further Discounts Editorial Board / Advisory Board members: 50%
Corresponding authors, affiliated to an EISZ member institution subscribing to the journal package of Akadémiai Kiadó: 100%
Subscription Information Gold Open Access
Purchase per Title  

Acta Chromatographica
Language English
Size A4
Year of
Foundation
1992
Publication
Programme
2021 Volume 33
Volumes
per Year
1
Issues
per Year
4
Founder Institute of Chemistry, University of Silesia
Founder's
Address
PL-40-007 Katowice, Poland, Bankowa 12
Publisher Akadémiai Kiadó
Publisher's
Address
H-1117 Budapest, Hungary 1516 Budapest, PO Box 245.
Responsible
Publisher
Chief Executive Officer, Akadémiai Kiadó
ISSN 0236-6290 (Print)
ISSN 2083-5736 (Online)

Monthly Content Usage

Abstract Views Full Text Views PDF Downloads
Feb 2021 0 40 14
Mar 2021 0 22 5
Apr 2021 0 25 6
May 2021 0 35 44
Jun 2021 0 19 22
Jul 2021 0 11 12
Aug 2021 0 0 0