Abstract

Tramadol, a strong pain killer known for its addictive problems is either co-administrated or co-formulated with other analgesics or muscle relaxants. The power of fluorescence detection in HPLC is tested to resolve such mixtures in plasma matrix to reach the required sensitivity with simple sample treatment using just protein precipitation. The aim of this work was to develop an eco-friendly and sensitive HPLC method with fluorimetric detection for analysis of Tramadol in its two binary mixtures with Ibuprofen (mixture 1) and Chlorzoxazone (mixture 2) in two combined dosage forms and spiked plasma. Separation was done using a C18 column with mobile phase of acetonitrile and water (pH 3.5) in gradient elution and 1 mL/min flow rate. Detection was carried out with λ excitation/λ emission of 220 and 307 nm, respectively. The method was applied to detect the two binary mixtures in real plasma samples after invivo application to rats, to assure that the drugs’ metabolites do not affect the sensitivity or selectivity of the assay. Evaluation of greenness of the proposed method was done using semi-quantitative Eco‐Scale and new Green Analytical Procedure Index which showed that this method can be a greener alternative with higher sensitivity for analysis of both mixtures. The method (15 min-assay) was linear over concentrations of 0.1–10 μg/mL and 0.1–33 μg/mL in plasma. In addition, the proposed method was validated per ICH as well as FDA bioanalytical methods’ validation guidelines.

Abstract

Tramadol, a strong pain killer known for its addictive problems is either co-administrated or co-formulated with other analgesics or muscle relaxants. The power of fluorescence detection in HPLC is tested to resolve such mixtures in plasma matrix to reach the required sensitivity with simple sample treatment using just protein precipitation. The aim of this work was to develop an eco-friendly and sensitive HPLC method with fluorimetric detection for analysis of Tramadol in its two binary mixtures with Ibuprofen (mixture 1) and Chlorzoxazone (mixture 2) in two combined dosage forms and spiked plasma. Separation was done using a C18 column with mobile phase of acetonitrile and water (pH 3.5) in gradient elution and 1 mL/min flow rate. Detection was carried out with λ excitation/λ emission of 220 and 307 nm, respectively. The method was applied to detect the two binary mixtures in real plasma samples after invivo application to rats, to assure that the drugs’ metabolites do not affect the sensitivity or selectivity of the assay. Evaluation of greenness of the proposed method was done using semi-quantitative Eco‐Scale and new Green Analytical Procedure Index which showed that this method can be a greener alternative with higher sensitivity for analysis of both mixtures. The method (15 min-assay) was linear over concentrations of 0.1–10 μg/mL and 0.1–33 μg/mL in plasma. In addition, the proposed method was validated per ICH as well as FDA bioanalytical methods’ validation guidelines.

1 Introduction

Different pain killers together with muscle relaxants are commonly used either over the counter or prescribed by physicians to relief moderate to severe pain especially associated with different musculoskeletal conditions. Thus, several combinations of analgesics (with or without muscle relaxants) are manufactured by different companies and launched in several markets. However, a lot of these combinations are not FDA approved nor tested for their pharmacological interactions. Therefore, it is necessary to find suitable analytical tools for simultaneous determination of these co-administered drugs in the same dosage form and especially in plasma to be available in hands for testing their pharmacokinetic interactions whenever needed.

Among these combinations found in the Indian market are two combinations of tramadol hydrochloride (TM). The first combination is TM with chlorzoxazone (CL) a muscle relaxant. The second combination is with ibuprofen (IB), especially designed for osteoarthritis patients, where TM would be an auxiliary to IB non-steroidal anti-inflammatory action. These two mixtures of TM have been co-formulated in tablets dosage forms to be used generally as a pain reliever, in cases of moderate and severe pain associated with musculo-skeletal spasms or cancer diseases. Since, there are not enough studies for the pharmacological interaction between TM and IB or CL when combined in the same dosage forms. It was important to develop a pharmacokinetic tool to analyze the three drugs in plasma for further pharmacological testing of the combined drugs when required in the future.

TM is an opioid analgesic drug centrally acting that has been used since 1977 for the relief of strong physical pain.[1] It is official in Indian [2], British [3] and United States [4] Pharmacopoeias. Chemically it is (1RS,2RS)-2-[(dimethylamino)methyl]-1-(3-methoxyphenyl)cyclohexanol hydrochloride (Fig. 1a). IB, chemically is (2RS)-2-[4-(2-methylpropyl)phenyl]propanoic acid (Fig. 1b), belongs to the NSAIDs (non-steroidal anti-inflammatory drugs) and is also official in the three pharmacopeias.[2–4] CL is chemically, 5-Chloro-2-benzoxazolinone (Fig. 1c). It is a centrally acting muscle relaxant that is official in United States Pharmacopoeia [4].

Fig. 1.
Fig. 1.

Chemical structures of (a) Tramadol (TM); (b) Ibuprofen (IB); and (c) Chlorzoxazone (CL)

Citation: Acta Chromatographica 34, 2; 10.1556/1326.2021.00901

Binary combinations of TM with IB or CL in their dosage forms challenging ratios (1:8, TM: IB and 1:5, TM: CL) were estimated in the literature by very few reported analytical methods including spectrophotometry [5, 6], HPLC [1, 6, 7], HPTLC [7, 8] and spectrofluorimetry [9]. However, by searching the literature, there were not any methods reported for the determination of the two binary mixtures in biological samples. In a resource limited settings, highly sensitive sophisticated instruments as mass spectroscopy would be not available and costly. An HPLC with fluorescence detector is a good choice to provide high sensitivity, compared to commonly used UV detection, and low quantitation limit (QL) which are essential for plasma analysis.

Therefore, the aim of this work was to establish a fast and simple HPLC method with fluorescence detection, due to its high sensitivity, for the simultaneous determination of TM, IB and CL in bulk form, dosage forms and plasma. Validation was fulfilled in accordance with the ICH “International Conference on Harmonization” [10] and the “FDA” Food and Drug Administration [11].

2 Experimental

2.1 Chemicals and reagents

TM (99.0% purity), IB (99.9% purity) and CL (99.9% purity) were kindly provided by Fluka BioChemika (Switzerland), EIPICO Pharmaceutical Industries (Egypt) and GlaxoSmithKline Co. (Egypt), respectively. Ibudol® tablets containing 400 mg IB and 50 mg TM and MUZOX® tablets containing 250 mg CL and 50 mg TM (Stedman Pharmaceuticals Pvt Ltd., India) were purchased from the local Indian market. Acetonitrile HPLC-grade (Sigma-Aldrich Chemie GmbH, Switzerland), ortho-phosphoric acid (BDH Laboratory Suppliers, England) and double distilled water have been used.

2.2 Instrumentation

Chromatographic analysis was performed using Agilent 1260 device (USA) equipped with auto-injector, quaternary pump and fluorescence detector (G1321C). Agilent ChemStation Software also has been used for data analysis. A reversed phase Agilent Zorbax SB-C18 (150 × 4.6 mm) column thermostated at ambient temperature (25 °C) has been used for the chromatographic separation with mobile phase flow rate of 1 mL/min. The mobile phase used consisted of acetonitrile and water acidified with orthophosphoric acid to pH 3.5. Step wise gradient elution system has been used starting with 17 and 83% (v/v) acetonitrile and acidified water, respectively, from 0 to 2 min then switched to 70 and 30% (v/v) acetonitrile and acidified water, respectively, and this ratio was kept till the end (14 min). The fluorescence detector was set to 220/307 nm (λ excitation/λ emission) for determination of the three studied analytes (TM, IB and CL).

2.3 Methods

2.3.1 Standard stock solutions

Standard stock solutions of TM, IB and CL (1000 µg/mL) were prepared by dissolving 100 mg, accurately weighed, of the authentic material of each drug in 100 mL methanol. Stability of the three prepared stock solutions was 14 days when stored at 4 °C protected from light.

2.3.2 Calibration graphs for TM and IB/CL analysis in dosage forms

TM, IB and CL standards (1000 μg/mL) were diluted by methanol to working solutions of 200 μg/mL each. The calibration standards were prepared by transferring accurate micro-volumes (5–500 µL) of each of TM, IB and CL working solutions into three separate sets of 10 mL volumetric flasks and diluted to volume with methanol to obtain a final calibration range of 0.1–10 µg/mL for all drugs. Triplicates HPLC injections (20 μL) of each drug calibration solution were injected using the chromatographic conditions previously mentioned (Section 2.2). The obtained peak area means were plotted against their corresponding concentrations so the calibration curves and regression equations would be obtained.

2.3.3 Calibration graphs for TM and IB/CL analysis in plasma

A series of working standards for plasma calibration curve and quality control samples "QCs" were prepared. Five different volumes of each of TM, IB and CL standard stock solutions (1000 µg/mL) were transferred into three series of five 10 mL volumetric flasks separately and completed to mark with methanol to achieve a final working concentration range of (1–330 µg/mL). Aliquots of 250 μL plasma were spiked separately with 25 μL of each of the TM and IB or TM and CL working solutions (1–330 µg/mL) and 25 μL of IS solution, from working solution of 100 μg/mL, (CL as IS for TM/IB mixture and IB as IS for TM/CL mixture) followed by 500 μL of acetonitrile to achieve concentrations in the range of 0.1–33 μg/mL plasma for the three concerned drugs with fixed IS concentration of 10 μg/mL plasma in all samples. After vortex mixing for 2 min, centrifugation was done for 15 min in a cooling centrifuge (−4 °C) at 15,000 rpm. The supernatant was then filtered by a 0.45 μm syringe adapter. A 20 μL volume was then injected in triplicate and chromatographed. In addition, QCs were prepared in the same way, 0.1, 0.3, 10 and 30 µg/mL plasma, by spiking 25 μL of their respective working standards in blank plasma to be LLOQ “lower limit of quantification”, LQC “low QC”, MQC “mid QC” and HQC “high QC”, respectively. The ratios of peak areas of analyte/IS against corresponding drug concentrations were plotted for calibration graphs construction.

2.3.4 Analysis of synthetic mixtures

Accurate volumes of standard solutions of TM, IB or CL were transferred into a set of five 10 mL volumetric flasks then dissolved to volume with methanol. Portions of 20 μL of each mixture in triplicates were chromatographed and the % recoveries were calculated.

2.3.5 Analysis of TM and IB/CL dosage forms

Five Ibudol® tablets were weighed and powdered. Into a 25 mL volumetric flask, a powder weight equivalent to 25 mg IB and 3.125 mg TM was transferred using 15 mL methanol. Five MUZOX® tablets were also weighed and powdered. Into another 25 mL volumetric flask, accurate powder weight equivalent to 25 mg CL and 5 mg TM was transferred. After 15 min sonication, the flasks were completed to volume with methanol and filtered using filter paper (Whatman). Aliquots of 80 and 50 µL from each of the Ibudol® tablet sample solution (mixture 1) and the MUZOX® tablet sample solution (mixture 2), respectively, were transferred separately into two 10 mL volumetric flasks and diluted to mark with methanol. The concentrations achieved after dilution were 8 μg/mL of IB and 1 μg/mL of TM (mixture 1) and 5 μg/mL CL and 1 μg/mL TM (mixture 2). Aliquots of 20 μL were injected and chromatographed as previously. The percentage recoveries were calculated using the previously computed regression equation (section 2.3.2).

2.3.6 In vivo application

Two groups each of six male Wistar rats weighing 200–250 gm were used for the in vivo study. Animals were housed under controlled conditions (25 ± 10 °C, RH 55 ± 10%). They were allowed to adapt to the housing environment for at least 1 week before the study. Diet was prohibited for 12 h before the experiment but water was freely available. Group (1) was treated with 20 mg/kg TM [12] and 17.8 mg/Kg IB [13] and group (2) was treated with 20 mg/kg TM [12] and 15 mg/Kg CL [14]. All drugs were orally administered once, and then six blood samples were collected from each group by retro-orbital bleeding at 3 specific time intervals for each drug in its mixture, which are tmax, taverage (average time interval between tmax and t1/2) and t1/2 of each drug. The blood samples were immediately collected to K2-EDTA tubes and centrifuged at 4000 rpm (10 min) to obtain plasma. The plasma samples were then stored at −80 °C until analysis time.

3 Results and discussion

3.1 Chromatographic method development and optimization

Several factors affected the separation and resolution of TM in its binary mixtures with IB and CL. All these factors were optimized independently to achieve the best chromatographic system suitability parameters. For selection of the stationary phase, some columns have been tried such as C8 (250 × 4.6 mm), C18 (150 × 4.6 mm) and C18 (250 × 4.6 mm). However, the best separation in terms of sensitivity, resolution and baseline noise was achieved using the Zorbax SB-C18 (150 × 4.6 mm) column. Different mobile phases were tried in order to achieve the best separation between TM and IB and/or CL in a reasonable analysis time. Mobile phases of different organic modifiers in different ratios (methanol, ethanol or acetonitrile) were tried. Acetonitrile gave the best results in terms of separation and peak symmetry. A high acetonitrile ratio was essential to elute IB but that caused TM to co-elute with the solvent front. Several isocratic trials were attempted to try to reach a mobile phase ratio that separated both TM and IB within reasonable retention times and also away from early retention times were the plasma co-eluting peaks may interfere. However, none of these trials succeeded due to the huge difference in retention times of TM and IB, so the only solution was to go for gradient programming. After numerous trials to reach the optimum gradient system to achieve the required separation between the drugs of interest, the system of choice was to start with 17 and 83% (v/v) acetonitrile and acidified water, respectively, till 2 min then switched to 70 and 30% (v/v) acetonitrile and acidified water, respectively. Different buffers such as phosphate, acetate and formate buffers at different pH values (pH 3.0–7.0) were tried in the beginning of the study. Using buffers resulted in peaks tailing and distortion of the baseline. Also, pure water caused broadening of the peaks due to its neutral pH value so water with the addition of phosphoric acid to maintain the pH acidic (3.5) was used.

It was important to optimize the pH of acidified water so pH 3 to 9 was tried. The optimum pH was 3.5 as it gave the best separation and least tailing for TM and CL and also after optimizing the gradient system and flow rate, this pH was also optimum for IB separation and peak symmetry.

Flow rate was also optimized to enhance the resolution of peaks. The flow rate has been adjusted over the range of 0.5–2 mL/min. The optimum chosen flow rate was 1 mL/min, which gave the best separation in reasonable retention time. Flow rates below 1 mL/min caused prolonged retention time. Increasing the flow rates caused resolution and retention time to decrease. After optimization of chromatographic parameters, all system suitability parameters have been measured and compared to reference values as they are essential for validation of the analytical method (Table 1).

Table 1.

System suitability parameters of the HPLC system used

Mixture Analyte Retention time (Rt), min Capacity factor (k′) Selectivity (α) Resolution (Rs) Asymmetry (Af) Efficiency (plates/m)
Mixture 1 TM 4.92 2.04 2.96 49.01 0.92 25,978.79
IB 11.38 6.03 ------ ------ 1.11 103,488.39
Mixture 2 TM 4.92 2.04 2.04 27.6 0.91 25,809.41
CL 8.37 4.17 ------ ------ 1,10 68,952.58

k′ (2−10), α > 1, Rs > 2, Af (0.8–1.2) and plates.m−1 (> 2000)

3.2 Sample extraction optimization

Protein precipitation is the method of choice for extraction in several reports [15, 16] due to its simplicity and greenness. Protein precipitation does not involve extraction with amounts of organic solvents nor huge number of steps like liquid-liquid or solid-liquid extractions. Both methanol and acetonitrile were tried; but acetonitrile was a stronger protein precipitating agent. There was no need to add strong acids nor pH adjustment nor evaporating the solvent to dryness. The 2:1 acetonitrile/plasma ratio was sufficient to precipitate the proteins and obtain good recovery for all drugs with the required sensitivity without breaching the rules of green analysis.

3.3 Fluorescence sensitivity and internal standard selection

Fluorescence detection offers higher selectivity and sensitivity compared to UV. So this advantage was used in this study, as the three drugs have native fluorescent nature. Several excitation and emission wavelengths were tried. The best λ excitation was found to be 220 nm that gave highest fluorescence signals for the three drugs especially the minor component (TM) and the three drugs had acceptable corresponding emission at 307 nm. Regarding internal standard selection, it was difficult to find a fluorescent sensitive compound at an appropriate retention time, with minimum interference of the plasma matrix and high resolution under the proposed chromatographic conditions. Thus, CL was chosen as IS for TM/IB mixture and IB was chosen as an IS for TM/CL mixture.

3.4 Method Validation

The ICH [10] and the FDA [11] for bioanalytical method validation guidelines have been followed for validation.

3.4.1 Linearity and range validation

Linearity has been studied to determine the range in which the analytes’ responses are linear with their concentrations (n = 5). By plotting peak areas (for dosage form analysis) or peak areas/IS peak area (for plasma analysis) against drugs’ concentrations (Table 2), the graphs obtained showed linear relationships. The slopes, intercepts, and correlation coefficients obtained by the linear least squares regression treatment of the results together with the standard deviations of intercept (Sa), slope (Sb) and residuals (Sy/x), indicating the closeness of the points to the line, are presented for each compound in Table 2. All obtained results were within the acceptable limits indicating the validity of the method. The correlation coefficients (r  ≥  0.9990) together with the low F-values indicate the good linearity of the calibration graphs.

Table 2.

Parameters of regression for TM/IB and TM/CL determination by the proposed method

Parameter For dosage form analysis For plasma analysis
TM IB CL TM IB TM CL
Linearity range, (µg/mL) 0.1–10 0.1–10 0.1–10 0.1–33* 0.1–33* 0.1–33* 0.1–33*
QL (µg/mL) 0.10 0.10 0.10 0.10* 0.10* 0.10* 0.10*
DL (µg/mL) 0.03 0.03 0.03 0.03* 0.03* 0.03* 0.03*
Intercept 35.76 21.29 8.03 0.11 0.06 0.24 0.03
Slope 122.98 97.98 49.28 0.07 0.15 0.15 0.03
Correlation Coefficient, r 0.9995 0.9991 0.9996 0.9990 0.9993 0.9991 0.9995
Sa 12.14 12.65 3.92 0.03 0.05 0.06 9.21 x 10−3
Sb 2.34 2.43 0.75 1.89 x 10−3 3.24 x 10−3 3.75 x 10−3 5.50 x 10−4
Sy/x 18.40 19.17 5.94 0.05 0.08 0.10 0.01
F 2770.69 1620.59 4270.17 1561.72 2093.67 1669.57 2871.82
Significance F 1.51 x 10−5 3.37 x 10−5 7.90 x 10−6 3.57 x 10−5 2.30 x 10−5 3.23 x 10−5 1.43 x 10−5

“Sa” standard deviation of intercept, “Sb” standard deviation of slope and “Sy/x” standard deviation of residuals (*µg/mL plasma)

Generally, the proposed method offered a LLOQ 0.1 µg/mL in plasma for the three drugs. The reported Cmax of TM, CL and IB after oral administration in plasma is within the linearity range of the three drugs [17-19] and higher than the LLOQ of this assay. Thus, this method is enough sensitive for the simultaneous quantitation of TM and IB or TM and CL in routine pharmacokinetic studies. Representative chromatograms of plasma spiked with LLOQ levels are shown in Fig. 2.

Fig. 2.
Fig. 2.

HPLC with fluorescence detection chromatograms of TM and IB (mixture 1) and CL (IS) (a) and TM and CL (mixture 2) and IB (IS) (B) in extracted plasma spiked at the LLOQ levels

Citation: Acta Chromatographica 34, 2; 10.1556/1326.2021.00901

3.4.2 Detection Limit (DL) and Quantitation Limit (QL) validation

Signal to noise ratio “S/N” method was used to calculate DL and QL where the S/N ratio is 3/1 and 10/1 for DL and QL, respectively. The DL/QL values of the proposed HPLC method are proposed in Table 2.

3.4.3 Accuracy and precision validation

In order to assess the accuracy and precision of the proposed method in separating and resolving the two studied binary mixtures in different ratios. Five different synthetic mixtures were chosen and prepared as mentioned above in section 2.3.4. including the dosage form ratio of the two mixtures and analysed. The mean percentage recoveries, the relative error values and relative standard deviation values indicate excellent accuracy and precision (Table 3) and proved no matter the concentration of each drug relative to the other, the method separate the drugs accurately and precisely.

Table 3.

Precision and accuracy assessment of the proposed method for determination of TM/IB and TM/CL in synthetic mixtures

Mixture 1 Mixture 2
TM IB TM IB TM CL TM CL
Concentration (µg/mL) Mean % Recovery ± %RSD %Er Mean % Recovery ± %RSD %Er Concentration (µg/mL) Mean % Recovery ± %RSD %Er Mean % Recovery ± %RSD a %Er
Results of accuracy and Intra-day precision, (n=5)
1.00 8.00 99.20 ± 1.30 −0.80 100.97 ± 0.99 0.97 1.00 5.00 101.07 ± 1.50 1.07 98.10 ± 1.20 −1.90
0.10 2.00 100.30 ± 1.20 0.30 100.50 ± 1.79 0.50 0.10 2.00 100.77 ± 1.00 0.77 100.04 ± 1.99 0.04
3.50 10.00 100.05 ± 1.94 0.05 99.99 ± 1.02 −0.01 3.50 10.00 101.09 ± 0.89 1.09 100.75 ± 1.80 0.75
2.00 0.10 98.20 ± 1.00 −1.80 100.19 ± 0.90 0.19 2.00 0.10 99.87 ± 0.99 −0.13 99.05 ± 1.09 −0.95
8.00 0.50 100.55 ± 1.25 0.55 100.50 ± 1.99 0.50 5.00 0.50 99.90 ± 0.87 −0.10 100.65 ± 0.60 0.65
Grand mean 99.66 ± 1.34 −0.34 100.43 ± 1.34 0.43 Grand mean 100.54 ± 1.05 0.54 99.72 ± 1.34 −0.28
Results of accuracy and Inter-day precision, (n=5)
1.00 8.00 98.22 ± 1.80 −1.78 99.59 ± 0.90 −0.41 1.00 5.00 100.99 ± 1.00 0.99 100.40 ± 1.74 0.40
0.10 2.00 99.10 ± 0.89 −0.90 99.67 ± 1.87 −0.33 0.10 2.00 99.91 ± 1.50 −0.09 99.55 ± 1.99 −0.45
3.50 10.00 101.30 ± 0.99 1.30 98.30 ± 1.20 −1.70 3.50 10.00 102.00 ± 1.99 2.00 99.90 ± 1.23 −0.10
2.00 0.10 100.44 ± 0.88 0.44 100.09 ± 0.85 0.09 2.00 00.10 100.50 ± 1.98 0.50 100.55 ± 1.83 0.55
8.00 0.50 99.99 ± 1.78 −0.01 99.50 ± 1.11 −0.50 5.00 00.50 101.35 ± 1.01 1.35 101.59 ± 0.99 1.59
Grand mean 99.81 ± 1.27 −0.19 99,43 ± 1.19 −0.57 Grand mean 100.95 ± 1.50 0.95 100.40 ± 1.56 0.40

•“%Er” percentage of relative error

•“%RSD” percentage of relative standard deviation

In plasma, validation of accuracy and precision was done at the four QC levels. Replicates (n = 6) of each concentration were analyzed on same day or on several days for intra- and inter-day evaluation of precision, respectively (Table 4).

Table 4.

Precision and accuracy assessment of the proposed method for determination of TM/IB and TM/CL in plasma

Concentration Theoretical concentration (µg/mL) Intra-day Inter-day
Calculated concentration (µg/mL) Mean % Recovery ± SD* %RSD %Er Calculated concentration (µg/mL) Mean % Recovery ± SD* %RSD %Er
TM
LLOQ 0.1 0.301 100.33 ± 1.00 1.00 0.33 0.305 101.67 ± 1.68 1.65 1.67
LQC 0.3 0.605 101.83 ± 1.35 1.33 1.83 0.609 101.50 ± 1.22 1.20 1.50
MQC 10.0 10.110 101.10 ± 1.25 1.24 1.10 10.255 102.55 ± 0.88 0.86 2.55
HQC 30.0 30.051 100.17 ± 1.29 1.29 0.17 30.005 100.02 ± 1.77 1.77 0.02
IB
LLOQ 0.1 0.318 102.25 ± 1.55 1.52 2.25 0.350 105.00 ± 1.89 1.80 5.00
LQC 0.3 0.656 102.55 ± 1.99 1.94 2.55 0.625 102.00 ± 1.25 1.26 2.00
MQC 10.0 5.355 103.04 ± 0.85 0.82 3.04 10.077 101.55 ± 0.97 0.96 1.55
HQC 30.0 30.089 100.07 ± 1.00 1.00 0.07 30.187 104.10 ± 1.34 1.29 4.10
CL
LLOQ 0.1 0.321 103.88 ± 1.90 1.83 3.88 0.334 100.50 ± 1.00 1.00 0.50
LQC 0.3 0.678 101.66 ± 0.99 0.97 1.66 0.622 102.60 ± 1.69 1.65 2.60
MQC 10.0 10.550 104.07 ± 0.23 0.22 4.07 10.902 101.99 ± 1.25 1.23 1.99
HQC 30.0 30.099 100.67 ± 1.50 1.49 0.67 30.561 100.44 ± 1.00 1.00 0.44

*standard deviation of six determinations

3.4.4 Recovery validation

Recoveries of TM, IB and CL spiked to plasma, at LQC and HQC levels, were calculated by comparing the results of the spiked samples to the standards’ drugs solutions. All extraction recoveries were between 96 and 99% and the mean recovery of IS of mixture 1 was 98% and of mixture 2 was 96.5%. These results indicate successful complete extraction just after simple protein precipitation pretreatment step.

3.4.5 Robustness validation

Robustness of the proposed method was assessed through analysis of synthetic mixtures with deliberate variation of selected chromatographic parameters including pH of the acidified water and its percentage, flow rate and temperature. All deliberate variations showed no influence on resolution or recoveries of the studied drugs (Table 5).

Table 5.

Robustness assessment for determination of TM/IB and TM/CL

Parameters tested HPLC–FD method*
TM IB CL
RSD % of peak areas Rt  ± SD RSD % of peak areas Rt ± SD RSD % of peak areas Rt ± SD
1) Mobile phase ratio [±1 % organic phase] 0.50 4.92 ± 2.0 × 10−2 1.50 11.38 ± 1.5 × 10−2 1.80 8.36 ± 2.2 × 10−2
2) Flow rate [1 ± 0.1 mL/min] 0.41 4.91 ± 2.5 × 10−2 0.80 11.38 ± 3.9 × 10−2 1.00 8.37 ± 4.5 × 10−2
3) Column temperature [25 oC ± 2 oC] 0.12 4.92 ± 3.4 × 10−2 1.50 11.36 ± 5.5 × 10−2 1.97 8.36 ± 5.5 × 10−2
4) pH of the aqueous phase [3.5 ± 0.5] 0.90 4.93 ± 3.6 × 10−2 0.65 11.39 ± 3.1 × 10−2 1.55 8.37 ± 1.54 × 10−2

* Mean of 3 concentrations levels 0.5, 5 and 10 μg/mL of TM, IB and CL

3.4.6 Selectivity validation

Selectivity of the proposed HPLC method for TM and IB/CL combined tablets analysis, was previously assessed in the results where the percentage relative errors < 2% proves the selectivity well. The absence of interference from the tablets’ excipients, also demonstrates the selectivity of the method (Table 6).

Table 6.

Assay results for the determination of TM, IB and CL in their drug products using the proposed method

Mixture 1 (TM/IB) Ibudol® tablets % Found ± SDa, n = 5
TM IB
Proposed HPLC method Comparison HPLC-UV method [6] Proposed HPLC method Comparison HPLC-UV method [6]
99.71 ± 0.87 100.78 ± 0.54 100.39 ± 0.91 100.51 ± 0.49
Students' t-test (t) * 0.07 0.82
Variance ratio F-test (F) * 2.62 3.47
Mixture 2 (TM/CL) MUZOX® tablets % Found ± SDa, n = 5
TM CL
Proposed HPLC method Comparison HPLC-UV method [1] Proposed HPLC method Comparison HPLC-UV method [1]
101.13 ± 0.83 101.56 ± 0.96 100.72 ± 1.18 100.48 ± 1.06
Students' t-test (t) * 0.58 0.81
Variance ratio F-test (F) * 0.20 0.13

* Theoretical values, at 95% confidence limit, of t and F are 2.31 and 6.39, respectively

The HPLC peak purity of the TM, IB and CL was also checked. The purity angle was within the purity threshold limit for all drugs. In addition, the emission spectra of TM, IB and CL were recorded at different time intervals across their peaks indicates the purity of the peaks in their dosage forms (Fig. 3) and also in plasma samples (Fig. 4).

Fig. 3.
Fig. 3.

Emission spectra illustrating peak purity of (a) TM and (b) IB (mixture 1) and (c) TM and (d) CL each is obtained from their tablets solutions for the HPLC method

Citation: Acta Chromatographica 34, 2; 10.1556/1326.2021.00901

Fig. 4.
Fig. 4.

Emission spectra illustrating peak purity of (a) TM and (b) IB (mixture 1) and (c) TM and (d) CL in plasma samples for the HPLC method

Citation: Acta Chromatographica 34, 2; 10.1556/1326.2021.00901

In plasma samples, no evidence of interfering peaks was noticed throughout the analysis. Chromatograms showing blank plasma and plasma spiked only with the ISs of the two mixtures are shown in Fig. 5a and b, respectively where no interference is noticed.

Fig. 5.
Fig. 5.

HPLC with fluorescence detection chromatograms for treated blank plasma (A) and treated blank with IS plasma (B)

Citation: Acta Chromatographica 34, 2; 10.1556/1326.2021.00901

3.4.7 Stability

The stability results of TM, IB and CL were obtained at LQC and HQC (n = 6) under freeze–thaw, postpreparative, short-term and long-term stability conditions in Table 7. The stock solutions and working standard of TM, IB and CL were stable for 14 days at refrigerator temperature and for at least 6 h at ambient temperature. The RSD% not exceeding 4% and Er% not exceeding 10% indicate good stability of the drugs under all the studied conditions. Further forced degradation studies will be conducted by the authors in a future study to investigate the stability of the combinations under different conditions as done in their previous assays [20].

Table 7.

Stability tests of TM, IB and CL in plasma (n = 6)

Stability TM IB CL
QC samples Mean % Recovery ± %RSD %Er Mean % Recovery ± %RSD %Er Mean % Recovery ± %RSD %Er
Short-term room temperature (6 h) LQC 103.02 ± 1.90 3.02 99.00 ± 1.67 −1.00 102.43 ± 0.70 2.43
HQC 102.22 ± 0.90 2.22 103.89 ± 1.70 3.89 100.50 ± 1.20 0.50
Freeze-thaw (3 cycles, at −70 °C) LQC 102.40 ± 1.77 2.40 100.34 ± 1.09 0.34 99.00 ± 0.71 −1.00
HQC 101.09 ± 1.50 1.09 101.99 ± 1.20 1.99 101.76 ± 0.50 1.76
Post preparative 5 °C for 24 h LQC 101.56 ± 1.45 1.56 98.50 ± 0.90 −1.50 100.40 ± 1.99 0.40
HQC 100.21 ± 0.87 0.21 97.20 ± 0.99 −2.80 99.01 ± 1.98 −0.99
Long-term 45 days at −70 °C LQC 103.44 ± 1.12 3.44 100.60 ± 0.92 0.60 97.54 ± 0.98 −2.46
HQC 100.46 ± 1.00 0.46 101.16 ± 1.66 1.16 100.09 ± 0.43 0.09

3.5 Application for determination of TM and IB/CL in their combined tablets dosage form

The proposed method was implemented for the analysis of TM in its co-formulated tablets with either IB or CL. Figure 6 shows the HPLC chromatograms, of the assayed pharmaceutical preparations with no interferences. The obtained results were not only in agreement with the labeled claims but also compared with those of the comparison methods [1, 6] (Table 6).

Fig. 6.
Fig. 6.

HPLC with fluorescence detection chromatograms of (a) Ibudol® tablet sample solution containing 1 μg/mL of TM and 8 μg/mL of IB (mixture 1) and (b) MUZOX® tablet sample solution containing 1 μg/mL of TM and 5 μg/mL CL (mixture 2)

Citation: Acta Chromatographica 34, 2; 10.1556/1326.2021.00901

3.6 Application for determination of TM and IB/CL in plasma samples

The developed method was applied successfully for determination of TM and IB/CL in plasma with only simple protein precipitation step with excellent correlation coefficients for the calibration curves (0.1–33 µg/mL plasma) and very low LLOD below the reported Cmax of the three studied drugs.

3.7 In vivo application

The applicability of the method to real clinical practice was assessed by therapeutic drug monitoring of the two binary mixtures in plasma samples from laboratory rats orally treated with the concerned drugs. The proposed method was applied for the determination of TM and IB or CL in plasma after a single oral dose of TM in combination with IB or CL. Real plasma samples were obtained at different intervals as collected in Table 8. At the day of analysis, calibration standards and QC samples were analyzed in parallel with the real plasma samples. Figure 7 shows representative chromatogram of real rat plasma samples with no interference from any metabolites that may have not appeared in spiked plasma samples. The peak shape and chromatographic resolution of the analytes were identical to those obtained from spiked plasma samples. The obtained results in Table 8 and Fig. 7 ensures the ability of the method to assay the drugs after their administration with no interference and with the required selectivity and accuracy. All determined concentrations were within the proposed calibration range. In addition the Cmax of all drugs obtained either from group one (treated with TM/IB) or group two (treated with TM/CL) were in compliance with the previously reported pharmacokinetic studies for each single drug. [12–14, 21, 22].

Table 8.

Data regarding drug posology and the corresponding plasma concentration of TM, IB & CL collected from male Wistar rats after applying the developed method

Mixture Monitored Drug Dose (mg/Kg) (single oral dose) Time sample collection (min) Mean plasma concentration ± SD (µg/mL)
Mixture 1 (TM, IB)

(n = 6)
TM 20 Tmax (60) 1.15 ± 2.09
Taverage (180) 0.91 ± 1.98
t1/2 (300) 0.50 ± 2.50
IB 17.8 Tmax (32) 35.0 ± 1.50
Taverage (70) 25.56 ± 0.99
t1/2 (108) 11.67 ± 1.50
Mixture 2 (TM, CL)

(n = 6)
TM 20 Tmax (60) 1.18 ± 1.22
Taverage (180) 1.03 ± 2.00
t1/2 (300) 0.55 ± 2.66
CL 15 Tmax (20) 1.60 ± 0.87
Taverage (60) 1.09 ± 0.99
t1/2 (99) 0.77 ± 1.09
Fig. 7.
Fig. 7.

HPLC with fluorescence detection chromatograms of the two binary mixtures in extracted real rat plasma samples at 60 min after oral administration of 20 mg/kg TM and 17.8 mg/Kg IB (a) and 20 mg/kg TM and 15 mg/Kg CL (b), each spiked with its respective IS

Citation: Acta Chromatographica 34, 2; 10.1556/1326.2021.00901

3.8 Assessment of greenness of the method

Green chemistry gained huge concern recently as it can be perceived as a contribution of analysts to the sustainable development trend. The analytical methods’ green evaluation can be challenging due to the many factors to be considered in method development as eliminating or reducing the use of reagents and decreasing energy consumption and wastes. Some measures may be adopted to make greener approaches such as using safer and more degradable solvents and reducing hazardous wastes and number of steps in analytical procedures. However, in some cases, it is very difficult to reach full green chemistry in certain aspects. For example, one of the main challenges is the multi-step nature of most analytical processes. Also, Tobiszewski et al. [23] pointed out that the least green step in analytical process is sample preparation step which is sometimes would be difficult to simply or eliminate.

To assess greenness of analytical methods, in a qualitative approach, several Internet databases are useful such as the NEMI “National Environmental Methods Index”, or the Green Chemical Alternatives Wizard developed by Massachusetts Institute of Technology.

Van-Aken et al. [24] developed an interesting semi-quantitative approach to greenness evaluation of organic synthesis, that is called the Eco-Scale, this scales assumes that an ‘‘ideal’’ system, has a score of 100, which is 100% safe for both the analyst and the environment. Penalty points are used for parameters deviating from the ideal value are which lower the overall score. The higher the score, the greener the procedure is. In 2012, an Eco-Scale for evaluation of green analysis was developed in the same approach.[25] The Eco‐Scale result of the proposed HPLC method is 85 showing excellent greenness (Table 9). This high score is due to using only one organic modifier in a very low percentage which is acetonitrile, a low toxicity solvent, instead of the commonly used hazard combination solvents. In addition, the sample pretreatment was done by simple protein precipitation using also acetonitrile only.

Table 9.

Points of penalty according to the Eco Scale of the proposed HPLC method

Reagents Penalty points
Acetonitrile 4
Phosphoric acid 2
Instrumentation Penalty points
LC 1
Occupational hazard 0
Waste 6
Centrifuge 1
Sonicator 1
Sum of penalty points 15
Final score 85

Another new green assessment tool is the Green Analytical Procedure Index “GAPI”. It uses pentagrams to evaluate the environmental impact of each step in the procedure using colors of three: red, yellow and green to represent high, medium and low impacts, respectively. This visual presentation allows researchers to make judgments about conflicting green criteria and compare between different methods to choose the greenest one for a particular purpose.[9] Figure 8 shows the pentagrams of the proposed method having low environmental impact based on the used reagents and instrumentation.

Fig. 8.
Fig. 8.

Green assessment profile for the proposed HPLC with fluorescence detection method using the GAPI tool

Citation: Acta Chromatographica 34, 2; 10.1556/1326.2021.00901

4 Conclusion

This work proposes a robust and valid HPLC method with fluorescence detection for simultaneous analysis of TM and IB (mixture 1) and TM and CL (mixture 2) either in their combined tablets or in ] plasma samples. The HPLC method proposed showed many advantages, the main of which is the novelty of the method; as the literature lacks the presence of any method for analysis of these two mixtures in plasma matrix and/or using fluorimetric detector. It was proved that the suggested method can be applied for future pharmacokinetics of these drugs due to its low sensitivity which was indicated by lower DL and QL values and invivo rat application. In addition the proposed method is eco‐friendly and green. The greenness of the method has been assessed by two different scales which are the new tool GAPI and the analytical Eco‐Scale. These advantages make this method superior for the routine analysis of such combination in commercially available tablets and also plasma samples.

Conflicts of interest/Competing interests

Not applicable.

Funding

Not applicable.

References

  • 1.

    Dutta, M. ; Wakode, S. Anal. Chem. Indian J. 2018, 18, 127.

  • 2.

    The Indian Pharmacopoeia, Controller of publications, New Delhi, 1996.

  • 3.

    The British Pharmacopoeia Commission , The British Pharmacopoeia, TSO, London, United Kingdom, 2018.

  • 4.

    USP-NF The United States Pharmacopeia, 33rd ed.—The National Formulary, 28th ed. United States Pharmacopeial Convention, Inc, USA, 2011.

    • Search Google Scholar
    • Export Citation
  • 5.

    Puranik, M. ; Hirudkar, A. ; Wadher, S. ; Yeole, P. Indian J. Pharm. Sci. 2006, 68, 737.

  • 6.

    Thomas, A. ; Dumbre, N. ; Nanda, R. ; Kothapalli, L. ; Chaudhari, A. ; Deshpande, A. Chromatographia 2008, 68, 843.

  • 7.

    Ahmed, H.M. ; Elshamy, Y.S. ; Talaat, W. ; Labib, H.F. ; Belal, T.S. Microchem. J. 2020, 153, 104505.

  • 8.

    Gandhimathi, M. ; Ravi, T.K. JPC-J. Planar Chromat. 2008, 21, 305.

  • 9.

    Abdel Moneim, M.M. ; Hamdy, M.M.A. Luminescence 2020.

  • 10.

    ICH Guidelines Validation of Analytical Procedures: Text and Methodology Q2 (R1), International Conference on Harmonization, Geneva, Switzerland, 2005, pp 1112.

    • Search Google Scholar
    • Export Citation
  • 11.

    Food, U. , Administration, D. , Guidance for Industry, Bioanalytical Methods Validation, 2001.

  • 12.

    Zhang, H. ; Zhao, Y. ; Wang, X. ; Zhang, Q. Biopharm. Drug Dispos. 2014, 35, 525.

  • 13.

    Medina, J.R. ; Jung, H. ; Hurtado, M. ; Soria, O. ; López-Muñoz, F. J. Int. J. Pharm. Sci. Res. 2017, 8, 1.

  • 14.

    Wang, X. ; Hu, L. ; Tong, S. ; Zheng, Y. ; Ye, F. ; Lin, D. ; Lin, G. ; Zhang, X. ; Wu, H. Anal. Lett. 2010, 43, 2424.

  • 15.

    Lu, W. ; Zhao, S. ; Gong, M. ; Sun, L. ; Ding, L. J. Pharm. Anal. 2018, 8, 160.

  • 16.

    El-kimary, E.I. ; Khamis, E.F. ; Belal, S.F. ; Abdel Moneim, M.M. J. Chromatogr. B Biomed. Appl. 2018, 94, 1097.

  • 17.

    Ardakani, Y.H. ; Rouini, M.R. Biopharm. Drug Dispos. 2007, 28, 527.

  • 18.

    Prompila, N. ; Wittayalertpanya, S. ; Komolmit, P. J. Med. Assoc. Thai 2007, 90, 160.

  • 19.

    Rainsford, K. Inflammopharmacology 2009, 17, 275.

  • 20.

    Ragab, M.A. ; El Yazbi, F.A. ; Hassan, E.M. ; Khamis, E.F. ; Hamdy, M.M. Anal. Chem. Lett. 2018, 8, 331.

  • 21.

    Shin, D ; Lee, S.J. ; Ha, Y.M. ,;Choi, Y.S. ; Kim, J.W. ; Park, S.R ,; Park, M.K. Drug Des. Devel. Ther. 2017,11,135.

  • 22.

    Gola, S. ; Gupta, A. ; Keshri, G.K. ; Nath, M. ; Velpandian, T. J. Pharm. Biomed. Anal. 2016, 121, 114.

  • 23.

    Tobiszewski, M. ; Mechlińska, A. ; Zygmunt, B. ; Namieśnik, J. Trac-Trend Anal. Chem. 2009, 28, 943.

  • 24.

    Van Aken, K. ; Strekowski, L. ; Patiny, L. Beilstein J. Org. Chem. 2006, 2, 3.

  • 25.

    Gałuszka, A. ; Migaszewski, Z.M. ; Konieczka, P. ; Namieśnik . Trac-trend Anal. Chem. 2012, 37, 61.

  • 1.

    Dutta, M. ; Wakode, S. Anal. Chem. Indian J. 2018, 18, 127.

  • 2.

    The Indian Pharmacopoeia, Controller of publications, New Delhi, 1996.

  • 3.

    The British Pharmacopoeia Commission , The British Pharmacopoeia, TSO, London, United Kingdom, 2018.

  • 4.

    USP-NF The United States Pharmacopeia, 33rd ed.—The National Formulary, 28th ed. United States Pharmacopeial Convention, Inc, USA, 2011.

    • Search Google Scholar
    • Export Citation
  • 5.

    Puranik, M. ; Hirudkar, A. ; Wadher, S. ; Yeole, P. Indian J. Pharm. Sci. 2006, 68, 737.

  • 6.

    Thomas, A. ; Dumbre, N. ; Nanda, R. ; Kothapalli, L. ; Chaudhari, A. ; Deshpande, A. Chromatographia 2008, 68, 843.

  • 7.

    Ahmed, H.M. ; Elshamy, Y.S. ; Talaat, W. ; Labib, H.F. ; Belal, T.S. Microchem. J. 2020, 153, 104505.

  • 8.

    Gandhimathi, M. ; Ravi, T.K. JPC-J. Planar Chromat. 2008, 21, 305.

  • 9.

    Abdel Moneim, M.M. ; Hamdy, M.M.A. Luminescence 2020.

  • 10.

    ICH Guidelines Validation of Analytical Procedures: Text and Methodology Q2 (R1), International Conference on Harmonization, Geneva, Switzerland, 2005, pp 1112.

    • Search Google Scholar
    • Export Citation
  • 11.

    Food, U. , Administration, D. , Guidance for Industry, Bioanalytical Methods Validation, 2001.

  • 12.

    Zhang, H. ; Zhao, Y. ; Wang, X. ; Zhang, Q. Biopharm. Drug Dispos. 2014, 35, 525.

  • 13.

    Medina, J.R. ; Jung, H. ; Hurtado, M. ; Soria, O. ; López-Muñoz, F. J. Int. J. Pharm. Sci. Res. 2017, 8, 1.

  • 14.

    Wang, X. ; Hu, L. ; Tong, S. ; Zheng, Y. ; Ye, F. ; Lin, D. ; Lin, G. ; Zhang, X. ; Wu, H. Anal. Lett. 2010, 43, 2424.

  • 15.

    Lu, W. ; Zhao, S. ; Gong, M. ; Sun, L. ; Ding, L. J. Pharm. Anal. 2018, 8, 160.

  • 16.

    El-kimary, E.I. ; Khamis, E.F. ; Belal, S.F. ; Abdel Moneim, M.M. J. Chromatogr. B Biomed. Appl. 2018, 94, 1097.

  • 17.

    Ardakani, Y.H. ; Rouini, M.R. Biopharm. Drug Dispos. 2007, 28, 527.

  • 18.

    Prompila, N. ; Wittayalertpanya, S. ; Komolmit, P. J. Med. Assoc. Thai 2007, 90, 160.

  • 19.

    Rainsford, K. Inflammopharmacology 2009, 17, 275.

  • 20.

    Ragab, M.A. ; El Yazbi, F.A. ; Hassan, E.M. ; Khamis, E.F. ; Hamdy, M.M. Anal. Chem. Lett. 2018, 8, 331.

  • 21.

    Shin, D ; Lee, S.J. ; Ha, Y.M. ,;Choi, Y.S. ; Kim, J.W. ; Park, S.R ,; Park, M.K. Drug Des. Devel. Ther. 2017,11,135.

  • 22.

    Gola, S. ; Gupta, A. ; Keshri, G.K. ; Nath, M. ; Velpandian, T. J. Pharm. Biomed. Anal. 2016, 121, 114.

  • 23.

    Tobiszewski, M. ; Mechlińska, A. ; Zygmunt, B. ; Namieśnik, J. Trac-Trend Anal. Chem. 2009, 28, 943.

  • 24.

    Van Aken, K. ; Strekowski, L. ; Patiny, L. Beilstein J. Org. Chem. 2006, 2, 3.

  • 25.

    Gałuszka, A. ; Migaszewski, Z.M. ; Konieczka, P. ; Namieśnik . Trac-trend Anal. Chem. 2012, 37, 61.

  • Collapse
  • Expand

Senior editors

Editor(s)-in-Chief: Sajewicz, Mieczyslaw, University of Silesia, Katowice, Poland

Editors(s)

  • Danica Agbaba, University of Belgrade, Belgrade, Serbia
  • Łukasz Komsta, Medical University of Lublin, Lublin, Poland
  • Ivana Stanimirova-Daszykowska, University of Silesia, Katowice, Poland
  • Monika Waksmundzka-Hajnos, Medical University of Lublin, Lublin, Poland

Editorial Board

  • Ravi Bhushan, The Indian Institute of Technology, Roorkee, India
  • Jacek Bojarski, Jagiellonian University, Kraków, Poland
  • Bezhan Chankvetadze, State University of Tbilisi, Tbilisi, Georgia
  • Michał Daszykowski, University of Silesia, Katowice, Poland
  • Tadeusz H. Dzido, Medical University of Lublin, Lublin, Poland
  • Attila Felinger, University of Pécs, Pécs, Hungary
  • Kazimierz Glowniak, Medical University of Lublin, Lublin, Poland
  • Bronisław Glód, Siedlce University of Natural Sciences and Humanities, Siedlce, Poland
  • Anna Gumieniczek, Medical University of Lublin, Lublin, Poland
  • Urszula Hubicka, Jagiellonian University, Kraków, Poland
  • Krzysztof Kaczmarski, Rzeszow University of Technology, Rzeszów, Poland
  • Huba Kalász, Semmelweis University, Budapest, Hungary
  • Katarina Karljiković Rajić, University of Belgrade, Belgrade, Serbia
  • Imre Klebovich, Semmelweis University, Budapest, Hungary
  • Angelika Koch, Private Pharmacy, Hamburg, Germany
  • Piotr Kus, Univerity of Silesia, Katowice, Poland
  • Debby Mangelings, Free University of Brussels, Brussels, Belgium
  • Emil Mincsovics, Corvinus University of Budapest, Budapest, Hungary
  • Ágnes M. Móricz, Centre for Agricultural Research, Budapest, Hungary
  • Gertrud Morlock, Giessen University, Giessen, Germany
  • Anna Petruczynik, Medical University of Lublin, Lublin, Poland
  • Robert Skibiński, Medical University of Lublin, Lublin, Poland
  • Bernd Spangenberg, Offenburg University of Applied Sciences, Germany
  • Tomasz Tuzimski, Medical University of Lublin, Lublin, Poland
  • Adam Voelkel, Poznań University of Technology, Poznań, Poland
  • Beata Walczak, University of Silesia, Katowice, Poland
  • Wiesław Wasiak, Adam Mickiewicz University, Poznań, Poland
  • Igor G. Zenkevich, St. Petersburg State University, St. Petersburg, Russian Federation

 

SAJEWICZ, MIECZYSLAW
E-mail:mieczyslaw.sajewicz@us.edu.pl

Indexing and Abstracting Services:

  • Science Citation Index
  • Sci Search
  • Research Alert
  • Chemistry Citation Index and Current Content/Physical
  • Chemical and Earth Sciences
  • SCOPUS
  • GoogleScholar
  • Index Copernicus
  • CABI
  • CABELLS Journalytics

2023  
Web of Science  
Journal Impact Factor 1.7
Rank by Impact Factor Q3 (Chemistry, Analytical)
Journal Citation Indicator 0.43
Scopus  
CiteScore 4.0
CiteScore rank Q2 (General Chemistry)
SNIP 0.706
Scimago  
SJR index 0.344
SJR Q rank Q3

Acta Chromatographica
Publication Model Online only
Gold Open Access
Submission Fee none
Article Processing Charge 400 EUR/article
Regional discounts on country of the funding agency World Bank Lower-middle-income economies: 50%
World Bank Low-income economies: 100%
Further Discounts Editorial Board / Advisory Board members: 50%
Corresponding authors, affiliated to an EISZ member institution subscribing to the journal package of Akadémiai Kiadó: 100%
Subscription Information Gold Open Access
Purchase per Title  

Acta Chromatographica
Language English
Size A4
Year of
Foundation
1988
Volumes
per Year
1
Issues
per Year
4
Founder Institute of Chemistry, University of Silesia
Founder's
Address
PL-40-007 Katowice, Poland, Bankowa 12
Publisher Akadémiai Kiadó
Publisher's
Address
H-1117 Budapest, Hungary 1516 Budapest, PO Box 245.
Responsible
Publisher
Chief Executive Officer, Akadémiai Kiadó
ISSN 2083-5736 (Online)

Monthly Content Usage

Abstract Views Full Text Views PDF Downloads
May 2024 0 38 15
Jun 2024 0 67 21
Jul 2024 0 79 32
Aug 2024 0 59 17
Sep 2024 0 75 30
Oct 2024 0 210 28
Nov 2024 0 11 1